Oxi 3310
Operation
33
ba75804e08
10/2017
the calibration. The evaluation appears on the display and in the cali-
bration record.
Calibration in
water vapor
saturated air
(air calibration vessel)
For this calibration procedure, the
Comparison meas.
setting must be
set to
Off
in the
Calibration
menu.
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.
Display
Calibration record
Relative slope
+++
S = 0.8 ... 1.25
++
S = 0.7 ... 0.8
+
S = 0.6 ... 0.7
Error
Error
S < 0.6 or S > 1.25
Eliminate the error according to
chapter 6 W
1
Connect the D.O. sensor to the meter.
2
Put the D.O. sensor into the air calibration vessel.
3
Start the calibration with
<CAL>
.
The last calibration data (relative slope) is displayed.
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......