Calibration datasets
The Oxi 3210 administrates two sets of calibration data:
Sensors of different types can be calibrated separately from each other.
When one sensor type is calibrated, the calibration data of the other
type remain stored. The Oxi 3210 recognizes the type of the connected
sensor and automatically uses the correct calibration data.
Calibration procedures
The Oxi 3210 provides as the calibration procedure:
Calibration in water vapor-saturated air.
Use an OxiCal
®
air calibration vessel for calibration.
Stability control
(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically
activated.
Display calibration data
You can have the data of the last calibration displayed (see section
4.4.5).
Calibration evaluation
After calibration, the meter automatically evaluates the current status of
the calibration. The evaluation appears on the display and in the cali-
bration record.
Calibration in
water vapor
saturated air
(air calibration vessel)
Proceed as follows to calibrate the instrument:
Set 1 for the type, "CellOx":
– CellOx 325
Set 2 for the type, "DurOx":
– DurOx 325
Display
Calibration record
Relative slope
+++
S = 0.8 ... 1.25
++
S = 0.7 ... 0.8
+
S = 0.6 ... 0.7
Error
Error
S < 0.6 or S > 1.25
Eliminate the error according to
chapter 6 W
1
Connect the D.O. sensor to the meter.
Summary of Contents for WTW Oxi 3210
Page 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Page 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Page 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Page 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Page 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Page 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Page 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...
Page 51: ......