NiCaVis 705 IQ TS
Commissioning
19
ba77189e01
05/2017
Measuring location
Outlet
(permanently set)
Measurement location or application of
the sensor.
The possible measurement loca
-
tions are displayed depending on the
currently set measuring mode.
The sample matrix on which the cal
-
culation of the measured value is
based changes with the location of
the sensor.
Measuring range
-
Display of the measuring range
(see chapter 8 T
ECHNICAL
DATA
).
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission
-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Number of C-sensors
1 ... 3
Apart from the main sensor, up to 4
more sensors can be displayed.
The settings for the additional sensors
are done in the menus for the respec
-
tive sensors.
Zero adjustment
Factory
User
Setting of which zero adjustment is
used in the sensor.
Factory zero adjustment or
Zero adjustment carried out last
Signal smoothing
On
Off
Switching on/off the signal smoothing.
For details, see
Response time t90
.
Menu item
Settings
Explanations