NiCaVis 705 IQ TS
Commissioning
17
ba77189e01
05/2017
3.4.2 Sensor structure
The physical sensor is the sensor with the instrument or model designation,
NiCaVis 705 IQ TS. The physical sensor processes the measurement signal
(absorbance spectrum) and supplies up to 5 measurement results to the
IQ S
ENSOR
N
ET
. The measurement results are administrated like 5 sensors that
differ in hierarchy as follows:
Main sensor The sensor for nitrate measurement is the main sensor.
It registers on the system under the designation, NiCaVis 705 IQ TS and the
series number of the physical sensor.
Secondary sensor The sensor for the measurement of total suspended solids is the secondary
sensor.
It registers on the system under the designation, SolidVis 70x IQ and the series
number of the physical sensor.
Virtual sensors All other sensors for carbon measurement are virtual sensors.
You register on the system under the designation, NiCaVis 705 IQ TS and the
series number of the physical sensor.
Sensor overlapping
settings and
functions
Some settings and functions are sensor overlapping. Sensor overlapping set
-
tings are generally edited in the setting menu of the main sensor.
These are the most important sensor overlapping settings and functions:
Measuring location
Times of the measuring cycle
(
Meas. interval
,
Cleaning duration Air
,
Adjustment phase
etc.)
Signal smoothing