Commissioning
NiCaVis 705 IQ TS
18
ba77189e01
05/2017
3.4.3 Settings for the main sensor
Measuring cycle A measuring cycle consists of the cleaning procedure, the adjustment time for
the measuring system and the determination of the measured value. The follow
-
ing graphic demonstrates the relevant settings:
Fig. 3-6 Measuring cycle of the sensor
Carrying out
settings
Using <S>, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu (setting table) of the sensor. The
procedure is described in detail in your IQ S
ENSOR
N
ET
system operating
manual.
Default values are marked in bold.
Time
Adjustment time
Measurement
Measuring
cycle
Measuring interval
Cleaning
A change of the following settings will take some time (up to a few
minutes) until it can be seen in the measured value display:
Measuring mode
Measuring location
Number of N-sensors
Number of C-sensors
Further changes can only be made when the last changes are visi
-
ble in the measured value display.
Menu item
Settings
Explanations
Measuring mode
NO3-N
NO3
The measured parameter is displayed
in the selected citation form.
When changing the
Measuring mode
setting, always check the
Measuring
location
setting.