Settings
Multi 3320
70
ba77053e04
07/2018
10.2 Sensor-independent settings
10.2.1
System
To open the
Storage & config
menu, press the
<ENTER
_
>
key in the mea-
sured value display. After completing the settings, switch to the measured
value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
System
/
General
/
Language
Deutsch
English
(more)
Selects the menu language
System
/
General
/
Beep
On
Off
Switches on/off the beep on
keystroke
System
/
General
/
Illu-
mination
Auto
On
Off
Switches the display illumina-
tion on/off
System
/
General
/
Contrast
0 ...
50
...
100 %
Changes the display contrast
System
/
General
/
Switchoff time
10 min ...
1h
... 24 h
Adjusts the shut-off time
System
/
General
/
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperature values are dis-
played with the selected unit.
System
/
General
/
Stability control
On
Off
Switches on or off the auto-
matic stability control during
measurement
System
/
Interface
/
Baud rate
1200, 2400,
4800
, 9600,
19200
Baud rate of the data interface
System
/
Interface
/
Output format
ASCII
CSV
Output format for data trans-
mission (see section 12 T
System
/
Interface
/
Decimal separator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System
/
Interface
/
Output header
Output of a header for
Output
format
:
CSV