Settings
Multi 3320
66
ba77053e04
07/2018
10.1.8 Settings for conductivity measuring cells
The settings are made in the
Measurement
menu for the measured parameter,
conductivity. To open the settings, display the required measured parameter in
the measured value display and press the
<ENTER>
key. After completing the
settings, switch to the measured value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of the
last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to the
interface with
<F2
_
>
/
[USB out-
put]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
USB
-
Outputs the calibration records to the
interface.
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the measuring
cell (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.