18
ba15360d04
12/2021
Commissioning
K 800
Commissioning
Combination elec-
trode K 800
Conditioning, calibration, measurement
General informa-
tion
When operating the electrode ensure that
the filling opening (1) for the bridge electrolyte is open
the inner junctions (3) are covered with bridge electrolyte
no air bubbles are in the bridge electrolyte
the depth of immersion is within the optimum range:
Before measuring
1
Remove the storage vessel.
2
Fill ELY/BR/503/K bridge electrolyte into the filling opening so that
the inner junctions are covered with bridge electrolyte.
3
Rinse the combination electrode with deionized water.
4
Wipe the shaft using a clean paper towel.
Minimum depth of immer-
sion
The outer junction (4) must be covered
Maximum depth of im-
mersion
Approx. 1 cm below the fluid level (2) of the
bridge electrolyte
1
2
3
4
1
Condition every ion sensitive electrode in standard solution before
using it.
Summary of Contents for wtw K 800 DIN
Page 2: ...2 ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15360d04 12 2021 Technische Daten K 800...
Page 16: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...ba15360f04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 32: ...32 ba15360f04 12 2021 Caract ristiques techniques K 800...
Page 34: ...ba15360s04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 42: ...42 ba15360s04 12 2021 Especificaciones t cnicas K 800...
Page 43: ......