Aging
K 800
Aging
Please note that every (combination) electrode undergoes a natural aging
process. The response time increases and the slope decreases with the age
of the (combination) electrode. The following factors shorten the lifetime con-
siderably:
Incorrect storage
Special measuring conditions (e.g. organic solutions, frequent measure-
ment with high concentrations of interfering ions)
High temperatures
High changes in temperature
The warranty does not cover damage caused by measuring conditions and
mechanical damage.
Maintenance
Combination electrodes:
Refill any used up bridge electrolyte.
Install an exchange membrane cap.
NOTE
Never pull the connection cable of the electrode.
The cable might be damaged.
Storage
Between two
measurements
Condition the electrode in standard solution.
Overnight
Rinse the combination electrode with deionized water and dry it. Then put on
the storage vessel and store the electrode upright.
Longer than over-
night
Remove the bridge electrolyte, then rinse the combination electrode with
deionized water and dab it dry with a clean paper towel. Store the combi-
nation electrode in the storage vessel in a dry place. For the following
measurement, prepare the combination electrode as described in section
C
.
Cs
+
NH
4
+
Tl
+
H
+
Ag
+
Tris
+
Li
+
Na
+
0.3
6
6
10
1000
1000
2000
2000
Summary of Contents for wtw K 800 DIN
Page 2: ...2 ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15360d04 12 2021 Technische Daten K 800...
Page 16: ...ba15360d04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...ba15360f04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 32: ...32 ba15360f04 12 2021 Caract ristiques techniques K 800...
Page 34: ...ba15360s04 12 2021 K 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 42: ...42 ba15360s04 12 2021 Especificaciones t cnicas K 800...
Page 43: ......