Commissioning
ColorVis 705 IQ
22
ba77260e01
06/2021
Factor turbidity
0.1 ... 1.0 ... 2.0
This menu item is only available with:
Setting
Color
:
True (TC)
.
The turbidity of a test sample is part of the
sample matrix and affects the measured
value of online measurements.
The effect of the turbidity on the measured
value is eliminated in the laboratory by
measuring a filtrated sample.
The effect of the turbidity measured in the
lab can be entered in the sensor as the
Fac-
tor turbidity
to achieve a permanent
turbidity compensation.
Cal - # raw value 1
Value pairs of the user calibration for TSS measurement (see section
Note: Leave those values unchanged during the initial commissioning.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Main sensor
Information that there are further sensor
overlapping settings in the setting menu of
the relevant main sensor (
Meas. interval
,
Cleaning duration Air
,
Signal smoothing
etc.).
Save and quit
-
The system confirms the saving of the set
-
tings and the display switches to the next
higher level.
Quit
-
The display switches to the next higher
level without saving the new settings.
Menu item
Settings
Explanations