ColorVis 705 IQ
Commissioning
19
ba77260e01
06/2021
Default values are marked in bold.
Menu item
Settings
Explanations
Measuring mode
Hazen 340
Hazen 350
Hazen 390
Hazen 445
Hazen 455
Hazen 465
ISO 410
The measured parameter is displayed in the
selected citation form.
Hazen: Measurement based on DIN EN
ISO 6271-2:2015, at specified wavelength
ISO: Measurement based on DIN EN ISO
7887:2012-04 Method C, at 410 nm
Measuring location
Outlet
(permanently set)
Measurement location or application of the
sensor.
Measuring range
-
Display of the measuring range
).
Color
True (TC)
Apparent (AC)
TC: with turbidity compensation
(True Color)
With the
setting
True (TC)
, a
Factor turbid
-
ity
can be set (see below).
AC: without turbidity compensation
(Apparent Color)
Factor turbidity
0.1 ... 1.0 ... 2.0
This menu item is only available with:
Setting
Color
:
True (TC)
.
The turbidity of a test sample is part of the
sample matrix and affects the measured
value of online measurements.
The effect of the turbidity on the measured
value is eliminated in the laboratory by mea-
suring a filtrated sample.
The effect of the turbidity measured in the lab
can be entered in the sensor as the
Factor
turbidity
to achieve a permanent turbidity
compensation.
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commissioning.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Num. of Color-sen
-
sors
1 ... 2
Adjustable number of sensors for the mea
-
surement of the coloration.
The settings for the additional sensor are
done in the menu for the sensor.