3.4.4 Settings for ColorVis 705 IQ (virtual sensor)
Carrying out
settings
Using <S>, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu (setting table) of the sensor. The
procedure is described in detail in your IQ S
ENSOR
N
ET
system operating
manual.
Default values are marked in bold.
Save and quit
-
The system confirms the saving of the set
-
tings and the display switches to the next
higher level.
Quit
-
The display switches to the next higher level
without saving the new settings.
Menu item
Settings
Explanations
Menu item
Settings
Explanations
Measuring mode
Hazen 340
Hazen 350
Hazen 390
Hazen 445
Hazen 455
Hazen 465
ISO 410
The measured parameter is displayed in
the selected citation form.
Hazen: Measurement based on DIN EN
ISO 6271-2:2015, at specified wave
-
length
ISO: Measurement based on DIN EN
ISO 7887:2012-04 Method C, at 410 nm
Measuring range
-
Display of the measuring range
Color
True (TC)
Apparent (AC)
TC (True Color):
with turbidity compensation
For the
True (TC)
setting, a
Factor tur
-
bidity
can be set (see below).
AC (Apparent Color):
without turbidity compensation