pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103
pHix Link software /Main configuration
64
Redox scaling
Output scaling
Menu item
Option
Default
Description
4mA [mV]
-1000 - 800
-1000
mV value represented by
4mA
20mA [mV]
-800 - 1000
1000
mV value represented by
20mA
Damping [s]
0 - 3600
10
Set the time period for
averaging the signal.
The averaging is a running
average.
The
minimum span is 200
.
Examples:
mV scaling from -1000 - 1000.
If there is 12 mA , the mV will be:
mV = 1000 -
(
(1000 - (-1000))/(20mA-4mA)
)
*(20mA-12mA)