3.7 Flow range / level range

This menu is mainly used to set the span of the mA out-
put in relation to the calculated flow, but can also be used 
to check the flow calculation in relation to the measured 
level based on the formula for the 
selected primary device / flow calculation.
Simply enter the flow with the arrow keys and press EN-
TER, and MJK 713 calculates the corres-ponding level.
The mA output span is set by entering the flow value 
where the mA output should give out 20 mA. 
An example: If a 12” Palmer & Bowlus flume is
selected as primary device, the nominal maximum flow 
will be 881 GPM at a level of 9.1 in.
If the flow range value is set to 881 in this menu, MJK 
713 calculates the level, the mA output will give out 20 
mA at max. nominal flow and 4 / 0 mA at zero flow.
If the flow range value is changed to 400 GPM, the mA 
output will give out 20 mA at 400 GPM and upwards and 
4 / 0 mA at zero flow.

4.0 Set digital output parameters

Press ENTER for configuring digital outputs or press 
MENU for MAIN MENU 

4.1 Select digital outputs (1-5)

Configuring of the five digital outputs. Select with the ar-
row keys which digital output to program, confirm with 
ENTER. 
Digital output 5 (DO5) is output for an external counter. 

4.2 Select function for DO

For digital output 1-4 select with arrow keys between 
various functions: counter output, sampler, flow > 0%, 
flow high, flow low and alarm sensor error. The choice is 
confirmed with ENTER.

Counter output:

  After a preset quantity the output is 

activated to trig an external counter.

Sampler:

 

After a preset quantity the output is 
activated to trig a sampling device or a 
chemical dosing device.

Flow > 0%:

 

Signal to indicate that the flow is greater 
than 0 is applied, i.e. when measuring 
emergency stormflow.

Flow rate high:

 

Activated if flow exceeds a given flow 
rate.

Flow rate low:

 

Activated if flow drops below a given flow 
rate.

Sensor error:

 

Activated at sensor error. Factory default 
is 5 minutes without an approved sensor 
signal.

Output disabled:

  Output not used.

4.3 Alarm flow rate low

The limit for flow low is keyed in using the arrow keys. 
The span is 0 - 90 % of max. flow.

4.4 Alarm flow rate high

The limit for high flow is keyed in using the arrow keys. 
The span is 10 - 100 % of max. flow.

4.5 Set volume between pulses

The volume that passes between each time a pulse is 
sent to a DO is keyed in. The interval between the pulses 
is limited to min. 1 sec.

4.6 Digital output ON-time

The ON time for the DO is keyed inusing the arrow keys. 
Double arrow changes the value fast with 1 sec. jumps. 
Single arrow changes the value slowly with 0.1 sec. 
jumps. The range is 0.1 - 30 sec.

Summary of Contents for MJK 713

Page 1: ...1 Manual US 3 10 713 FLOW CONVERTER 1905 SW 833062 MJK Automation Blokken 9 DK 3460 Birker d Denmark Tel 45 45 56 06 56 www mjk com MJK 713 Open Channel Flowmeter...

Page 2: ...7 conform ment aux dispositions de Directive 2004 108 EC to which this declaration relates is in conformity with the following standard s or other normative document s EN60079 0 2006 EN60079 11 2007 E...

Page 3: ...h weirs 9 Measuring along a wall or other plane 10 Flow measurement with smaller flumes weirs 10 The spread of the ultrasonic signal 10 Measuring through a concrete deck 11 Measurement through a pipe...

Page 4: ...ction 24 3 4 2 Rectangular weir crest width 24 3 5 Q h linearization 24 3 5 1 Number of Q h points 24 3 5 2 Select decimal point for the flow values 24 3 5 3 Linearization height value 24 3 5 4 Linear...

Page 5: ...is a Xylem brand Safety instructions 1 Read this manual carefully 2 Be aware of the environment on the installation site Wear necessary protective equipment and follow all current safety regulations 3...

Page 6: ...eir or flume The flow converter has 3 different linearization systems depending on how the volume of water is mea sured preprogrammed formulas for different dimensions of the most common flumes and we...

Page 7: ...wall A full scale print of above drawing can be found at the end of this manual 2a To ensure a reliable and precise level measurement it is of vital importance that the ultrasonic sensor points down a...

Page 8: ...The ultrasonic sensor should be installed so that the bottom of the sensor is as close as possible to the max height of the head but not closer than the blocking distance Measuring range 0 1 0 3 Bloc...

Page 9: ...out side contraction The distance L should be between 4 to 5 times the nomi nal width b Measuring distances for V notch weirs The distance L should be between 3 to 4 times the height h of the notch 4...

Page 10: ...ined See figure 1 Figure 1 The signal spread in conjunction with the measurement distance The signal spread should be increased by 50 100 if the surface is not smooth Flow measurement with smaller flu...

Page 11: ...e to build up foam on the surface it can become necessary to measure through a pipe since the build up of foam rarely will occur inside the pipe Measurement through a pipe for foam protection Measurin...

Page 12: ...ed in the correct height over the bottom of the flume weir 1 in NPT thread Actual measurement Measuring range Measuring distances for Palmer Bowlus flumes Diameter D Distance L Diameter D Distance L 4...

Page 13: ...Denmark Tel 45 45 56 06 56 www mjk com Measuring distances for rectangular weirs with and without side contraction The distance L should be between 4 to 5 times the nomi nal width b Measuring distanc...

Page 14: ...e No 4 yellow No 5 black MJK 713 2500 Flowmeter Ultrasonic Sensor MJK 713 System Package 1 MJK 713 2500 with ultrasonic sensor MJK 713 System Package 3 MJK 713 3000 with hydrostatic level transmitter...

Page 15: ...ultrasonic sensor cable has 5 wires with both color code and number Number Color Designation 1 Brown Ultrasonic pulse 2 Red 3 Orange Temperature compensation 4 Yellow 5 Black Shield This wire is conn...

Page 16: ...ording to the terminal mark ings on the connection box PCB and on the MJK 713 respectively Take care not to block or squeeze the air pressure compensation tube Cutting the cable The cable is delivered...

Page 17: ...NU key the display will change to the next main menu Arrow keys The arrow keys are used for changing a current setting Pressing the arrow keys change the values up or down By pressing the up arrow key...

Page 18: ...E simultaneously and pressing ENTER while in level readout menu Initial startup MJK 713 will show the following display at power up MJK 713 initiates the quick setup sequence after app 30 sec The foll...

Page 19: ...ow you configure the flow converter Configuring the flow converter is accomplished in the configuration menus which in turn are divided into sev eral sub menus You gain access to the configuration men...

Page 20: ...access to indication of summed quanti ties F5 Totalizer Press once for Here the total quantity is indicated from when the value was last reset The quantity is updated every second and is indicated wi...

Page 21: ...registred with a precision of 5 min End of power loss is registred when the flowmeter is powered up The upper line in the display shows the time for starting the alarm The bottom line shows the time f...

Page 22: ...5 0 indicates non activated output whereas 1 indicates activated outputs Language Press MENU twice for Change the language with the arrow keys and confirm with ENTER Enter access code Press MENU three...

Page 23: ...ent Level indication and unit of elevation can be chosen as m cm mm ft in Select with the arrow keys and confirm with ENTER 1 5 Select unit volume The unit for volume can be chosen between m3 Gal CF S...

Page 24: ...keys and confirm with ENTER The selected number of Q h points decides how many times the menus 3 5 3 to 3 5 4 will appear 3 5 2 Select decimal point for the flow values Select the decimal point and 1...

Page 25: ...o program confirm with ENTER Digital output 5 DO5 is output for an external counter 4 2 Select function for DO For digital output 1 4 select with arrow keys between various functions counter output sa...

Page 26: ...t be exceeded before a DO is activated The format is as follows min ss The maximum delay is 99 minutes and 59 seconds 4 8 Digital output NC NO The relay function for digital output is keyed in as Nor...

Page 27: ...rate today Press 3 X Average flow rate 24 hr Totalizer key Press 1 X Totalizer 1 hr Press 2 X Totalizer today Press 3 X Totalizer last 24 hr In the last selection the totalizer values for the last 99...

Page 28: ...US 3 10 713 FLOW CONVERTER 1905 SW 833062 MJK Automation Blokken 9 DK 3460 Birker d Denmark Tel 45 45 56 06 56 www mjk com Configuration menus 1 0 2 0 3 0 1 1 1 2 2 1 1 1 4 2 1 2 1 3 1 6 1 5 2 2 3 1 3...

Page 29: ...NVERTER 1905 SW 833062 MJK Automation Blokken 9 DK 3460 Birker d Denmark Tel 45 45 56 06 56 www mjk com Configuration menus 4 0 5 0 4 1 4 2 4 5 4 6 4 3 4 4 4 7 4 8 3 4 3 5 3 6 3 4 1 3 5 1 3 6 1 3 4 2...

Page 30: ...nt Flow MGD Sensor distance 30 in 54 in Averaging of Level 1 sec Range for flow 1 000 MGD 10 00 MGD Flow Calculations Parshall Flume 6 range 10 in Palmer Bowlus Flume 10 range 10 in V notch 90 Rectang...

Page 31: ...c c e r s i t I s e l b a c g n i d n e t x e e l t t u h S c i n o s a r t l U r o s n e S n o i t i d n o C d e k c a r c r o d e r u o l o c s i m r o s n e s e h t f o t r a p k c a l b e h t s I...

Page 32: ...voltage 10 30 V DC 12 30 V DC for cable lengths above 315 ft Output Two wire 4 20 mA passive transmitter Cable 2 AWG 20 shielded oil resistant PUR insulation Cable length 39 ft Longer cable lengths av...

Page 33: ...t be necessary Below is shown which adjustments might be necessary to carry out in field Ultrasonic measuring circuit Gain With this potentiometer the signal strength of the received signal can be adj...

Page 34: ...ols for indication during programming and normal service Keyboard Keyboard with a brief instruction for programming the SONOLEV The lower part can be tipped up to show the terminal designations When S...

Page 35: ...e the UP DOWN keys to set the span and confirm with ENTER Setting of relay output 1 If the mA output is looped through relay 1 as shown on the connection diagram on page 14 System Package 2 MJK 713 wi...

Page 36: ...45 45 56 06 56 www mjk com When the level starts to rise Shuttle will also indicate the level of the mA output with the bar graph One segment is roughly 1 mA Mount the lid when all settings are made S...

Page 37: ...ontraction 0 75 2 920 2 ft ft w contraction 1 00 5 994 2 ft w contraction 1 25 10 47 3 ft w contraction 1 50 16 52 4 ft w contraction 2 00 33 91 5 ft w contraction 2 50 59 23 6 ft w contraction 3 00 9...

Page 38: ...www mjk com Dimension Max head Max nominal flow ft CFS 22 5 V notch 2 00 2 811 30 V notch 2 00 3 824 45 V notch 2 00 5 855 60 V notch 2 00 8 163 90 V notch 2 00 14 14 120 V notch 2 00 24 49 V notch w...

Page 39: ...39 Manual US 3 10 713 FLOW CONVERTER 1905 SW 833062 MJK Automation Blokken 9 DK 3460 Birker d Denmark Tel 45 45 56 06 56 www mjk com...

Page 40: ...nish law on product liability however the liability is reduced to coverage of our public liability insurance of products To the extent where nothing else follows in lines of invariable rules of law we...

Reviews: