Outputs
DIQ/S 281
ba76226e02
03/2023
120
6.7.1
Recorder
Function
The measured values of the linked sensor at the current output are set up as cur-
rent intensity in the
Recorder
application. The output of measured values is laid
down in the
Recorder type
,
Start value
and
End value
settings.
Settings
9
Using
<
>
and
<OK>
, highlight and confirm
Save and quit
.
The new settings are stored.
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
(sensor
dependent)
Minimum spacing:
≤
5 % of the measuring range
(sensor-dependent)
End value
Measured
variable
Main variable
Adjoining variable
Main variable
designates the actual
measured parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an
additional measured parameter (e.g.
temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the output
current (mA/s) in the case of erratic
changes of the input signal.
I -> UFL/OFL
Error
Current values outside the range
between
Start value
and
End value
are regarded as an error. The current
output reacts as specified under
Behavior at error
(see below).
Limitation
The current at the output is limited to
the
Start value
or
End value
.
Behavior at error Fixed current
value
In the case of system errors and
sensor errors, the current output
delivers the fixed current value that
was specified.
Possible values:
0 ... 21 mA.
Unchanged
The current at the output remains
unchanged.