DIQ/S 281
Settings/setup
91
ba76226e02
03/2023
Alarm links setting
table
figure 5-9
Set alarm link
5
Edit the setting table. The required operating steps are described in
detail in section 4.2 G
Menu item
Selection/Values
Explanations
Measured variable
Main variable
Adjoining variable
Main variable
designates the actual
measured parameter of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured parameter (e.g.
temperature).
Limit value
Lower limit
Upper limit
Type of the alarm event.
Lower limit
: An alarm is triggered if the
defined limit value is undercut.
Upper limit
: An alarm is triggered if the
defined limit value is exceeded.
Lower limit
/
Upper limit
within the measuring
range (sensor-
dependent)
Limit value for the alarm event
Hysteresis
0 - 10 % of the
measuring range
Hysteresis for the limit value
Designation
(max. 20 characters)
User-defined designation for easier
identification in the alarm message.