Outputs
DIQ/S 281
ba76226e02
03/2023
116
* With short cleaning intervals, the adjustable values for the
Cleaning duration
and
Adjustment time
are limited. The following values apply:
Example
Cleaning interval
Cleaning duration
Adjustment time
≤
10 min
max. 60 s
max. 120 s
≤
20 min
max. 180 s
max. 300 s
With this, the cleaning times are fixed. They only change when the
Reference time (h)
is changed.
The reference time and all further cleaning times relate to the date
and time of the system clock. How to set the system clock is
described in the system operating manual.
Setting
Result
Reference time (h)
:
Reference time
(min)
:
Interval unit
:
Cleaning interval
:
12
0
Hours (h)
8 h
Reference time: 12:00 hours
This specifies the following start times:
04:00, 12:00 and 20:00 hours