Example of automatic Data Logger mode

 

Example: automatic recording of channel 1 on internal memory every 2 minutes. 

 

 

Connect the temperature probe only on channel 1;

 

 

Access to the LOG SETTING P8.0 setup menu;

 

 

Enter the LOG TYPE P8.1 menu, press the button 

 and move with directional keys to MINUTES;

 

 

Use the directional keys to change the number that flashes on the display. Enter "2" and confirm with 

key 

.

 

 

Return to measure mode.

 

The icon 

 is on, which indicates that an automatic frequency Data Logger has been set.

 

Press 

 to start recording; the icon 

 starts flashing, indicating that the storage is in progress. The 

number next to the 

M+

 icon indicates how much data has been saved.

 

 

Press again 

 to end the recording and the icon 

 returns fixed on the display.

 

 

Note

: if the automatic switch off is activated, the instrument will turn off after 20 minutes.  

 

For a longer recording is necessary to deactivate the auto-switch off  P9.9.

 

 

 

Instrument configuration menu

 

 

 

 

 

In measure mode, press button 

to access the SETUP menu;

 

 

Press the key 

 or 

 to move to SETTINGS menu P9.0;

 

 

Press the key  

 to access the menu;

 

 

Move with keys 

 and

 to select the program to access.

 

The table below shows the setup menu structure for the instrument, for each program, there are the options 
that the user can choose and the default value:

 

Configuration Settings menu

 

Program

 

Description

 

Options

 

Factory Default Settings

 

P9.1

 

TEMPERATURE UNIT

 

°C - °F

 

°C

 

P9.2

 

DATE AND TIME

 

---

 

---

 

P9.3

 

BACKLIGHT MODE

 

INDOOR / OUTDOOR / AUTO

 

AUTO

 

P9.4

 

BRIGHTNESS

 

LOW / MED / HIGH

 

MED

 

P9.5

 

SLEEP MODE

 

2 MIN / 5 MIN / OFF

 

2 MIN

 

P9.7

 

VIEW DIFFERENCE

 

YES / NO

 

NO

 

P9.8

 

RESET SETTINGS

 

YES / NO

 

NO

 

P9.9

 

AUTO OFF

 

YES / NO

 

YES

 

 

P9.1 Temperature unit

 

Access this setup menu to select the temperature unit to use: (°C - °F).

 

Note

: by setting a different unit of measurement, the offset will be convert to the new measure unit. The 

offset correction is shown when the instrument is turned on and in menu P7.3.

 

 

P9.2 Date and time setting

 

Access this setup menu to update the date and time of the device. Use the directional keys to change the 

year, confirm with button 

 and repeat the same operation for month, day, hours and minutes.

 

 
 
 

Summary of Contents for Temp70 PT100 vio

Page 1: ...INSTRUCTIONS MANUAL Temp70 PT100...

Page 2: ...1 UM Termometri Temp70 PT100 Vio ENG rev 0 05 04 2021...

Page 3: ...es 6 Datasheet 6 Instrument description 7 Display 7 Keypad 7 LED 8 Setup 8 Supplied components 8 Start up 8 Switching On Off 8 Batteries replacement 8 Instrument transportation 9 Key Functions 9 Input...

Page 4: ...100 Vio ENG rev 0 05 04 2021 Warranty 18 Warranty period and limitations 18 Disposal of electrical devices 18 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax 39059653...

Page 5: ...ided ATTENTION for a dangerous situation with reduced risk which can cause material damaged data loss or minor or medium sized accidents if not avoided WARNING for important information about the prod...

Page 6: ...rs or in environments with conditions different from those mentioned in this manual Device maintenance If used correctly and in a suitable environment the instrument does not require maintenance proce...

Page 7: ...and automatic with integrated sensor Sleep mode Yes Auto switch off Yes after 20 minutes deactivable LED indication Yes Reading stability indicator Yes adjustable Low Med High Stability filter HOLD Ye...

Page 8: ...ing function MAX MIN if active OK Setup Key to confirm the values and enter into the setup menu from the meausure mode Up directional Key Save data for scrolling through menus and parameters Press fas...

Page 9: ...m by pressing the button The display initially activates all segments and then appears model and firmware of the instrument information about the calibration of the channel 1 and the channel 2 screen...

Page 10: ...different views while the two probes are connected channel 1 channel 2 and both channels simultaneously Into the setup menus press to return to measure mode Short To scroll and modify the values of th...

Page 11: ...ay FIXED Automatic Data Logger set INTERMITTENT Automatic Data Logger in operation Instrument connected to the DataLink software Device operation On the left side of the display through a string of di...

Page 12: ...ibration data P7 4 Select resolution P7 5 Stability filter P7 6 Reset settings P8 0 LOG SETTINGS P8 1 Log type P8 2 Clear data P9 0 SETTINGS P9 1 Temperature unit P9 2 Date and time P9 3 Backlight mod...

Page 13: ...resolution for the temperature reading 0 1 0 01 Factory Default settings Note changing the temperature resolution the temperature resolution for both channels will change too P7 5 Stability filter To...

Page 14: ...uisition mode MANUAL The data is acquired only when the user presses the button HOURS MINUTES Set a frequency range of automatic data acquisition With the direction keys move from MANUAL to HOURS or M...

Page 15: ...configuration menu In measure mode press button to access the SETUP menu Press the key or to move to SETTINGS menu P9 0 Press the key to access the menu Move with keys and to select the program to ac...

Page 16: ...ly saving battery consumption Note To exit from the Sleep mode and return the display to normal brightness press ANY button Once the display brightness is activated the buttons reacquire their functio...

Page 17: ...screen on the display a text string will flash MIN MAX REC 1 probe connected REC both probes connected Press the key once to check the maximum value reached up to that moment press it another time to...

Page 18: ...h the key or and confirm with the key In the middle of the display the correct temperature is displayed while the partial value of the calibration is displayed below it The adjustment changes the enti...

Page 19: ...acturer will repair or replace defective components This warranty is valid only and exclusively on the electronic parts of the device and does not apply if the product has been damaged used incorrectl...

Reviews: