XS Instruments Temp70 PT100 vio Instruction Manual Download Page 11

 

10  

 

UM Termometri Temp70 PT100 Vio ENG rev.0 05.04.2021

 

 

Display symbols and icons 

 

 

Symbol

 

Description

 

 

Type of probe connected

 

 

Error in measurement

 

 

Battery charge indication

 

 

Measurement stability indicator

 

 

The bars scroll if the measurement is not stable

 

 

Number of data stored in Data Logger mode on instrumental memory

 

 

Press the directional keys to change the parameter or value on the display

 

 

FIXED: Automatic Data Logger set

 

INTERMITTENT: Automatic Data Logger in operation

 

 

Instrument connected to the D software

 

 

 

Device operation

 

 

On the left side of the display, through a string of different colours, it is always indicated how the instrument 
is located

.

 

String

 

Meaning

 

 

The instrument is in measure mode.

 

 

The instrument is in the Recall phase of the recorded MAX or 
MIN value.

 

The function HOLD is active.

 

Recall of the saved data.

 

 

The user is in the setup mode.

 

The configuration menus can concern the characteristics of the 
parameters or the general setting of the instrument

 

 

 
 
 

M+

 

Summary of Contents for Temp70 PT100 vio

Page 1: ...INSTRUCTIONS MANUAL Temp70 PT100...

Page 2: ...1 UM Termometri Temp70 PT100 Vio ENG rev 0 05 04 2021...

Page 3: ...es 6 Datasheet 6 Instrument description 7 Display 7 Keypad 7 LED 8 Setup 8 Supplied components 8 Start up 8 Switching On Off 8 Batteries replacement 8 Instrument transportation 9 Key Functions 9 Input...

Page 4: ...100 Vio ENG rev 0 05 04 2021 Warranty 18 Warranty period and limitations 18 Disposal of electrical devices 18 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax 39059653...

Page 5: ...ided ATTENTION for a dangerous situation with reduced risk which can cause material damaged data loss or minor or medium sized accidents if not avoided WARNING for important information about the prod...

Page 6: ...rs or in environments with conditions different from those mentioned in this manual Device maintenance If used correctly and in a suitable environment the instrument does not require maintenance proce...

Page 7: ...and automatic with integrated sensor Sleep mode Yes Auto switch off Yes after 20 minutes deactivable LED indication Yes Reading stability indicator Yes adjustable Low Med High Stability filter HOLD Ye...

Page 8: ...ing function MAX MIN if active OK Setup Key to confirm the values and enter into the setup menu from the meausure mode Up directional Key Save data for scrolling through menus and parameters Press fas...

Page 9: ...m by pressing the button The display initially activates all segments and then appears model and firmware of the instrument information about the calibration of the channel 1 and the channel 2 screen...

Page 10: ...different views while the two probes are connected channel 1 channel 2 and both channels simultaneously Into the setup menus press to return to measure mode Short To scroll and modify the values of th...

Page 11: ...ay FIXED Automatic Data Logger set INTERMITTENT Automatic Data Logger in operation Instrument connected to the DataLink software Device operation On the left side of the display through a string of di...

Page 12: ...ibration data P7 4 Select resolution P7 5 Stability filter P7 6 Reset settings P8 0 LOG SETTINGS P8 1 Log type P8 2 Clear data P9 0 SETTINGS P9 1 Temperature unit P9 2 Date and time P9 3 Backlight mod...

Page 13: ...resolution for the temperature reading 0 1 0 01 Factory Default settings Note changing the temperature resolution the temperature resolution for both channels will change too P7 5 Stability filter To...

Page 14: ...uisition mode MANUAL The data is acquired only when the user presses the button HOURS MINUTES Set a frequency range of automatic data acquisition With the direction keys move from MANUAL to HOURS or M...

Page 15: ...configuration menu In measure mode press button to access the SETUP menu Press the key or to move to SETTINGS menu P9 0 Press the key to access the menu Move with keys and to select the program to ac...

Page 16: ...ly saving battery consumption Note To exit from the Sleep mode and return the display to normal brightness press ANY button Once the display brightness is activated the buttons reacquire their functio...

Page 17: ...screen on the display a text string will flash MIN MAX REC 1 probe connected REC both probes connected Press the key once to check the maximum value reached up to that moment press it another time to...

Page 18: ...h the key or and confirm with the key In the middle of the display the correct temperature is displayed while the partial value of the calibration is displayed below it The adjustment changes the enti...

Page 19: ...acturer will repair or replace defective components This warranty is valid only and exclusively on the electronic parts of the device and does not apply if the product has been damaged used incorrectl...

Reviews: