Waterproof tester series 

7

 

INSTRUCTIONS FOR pH5 / Cond5 / PC5 

POWER ON 

Press   key once, the meter will switch On and performs a test of 3 colors backlight display and then show 
all the segments active for 2 sec.  
Then it will display the following: 

 

 

 

 

 

 

 

 

 

   

  

SETUP MENU for pH5 / Cond5 / PC5 

1.

 

With meter switched off, press and hold 

CAL

 key and press   key once. 

2.

 

The meter will switch on with all the segments active, release 

CAL

 key, the meter will go into the 

Setup Menu (green backlight during setup). 

3.

 

Press   key to select the parameter to be changed between: 
 

 

 

Function 

Primary display 

Secondary display 

Default value 

Temperature unit (°C/°F) 

t.U 

°C 

Reference temperature for conductivity 

trE 

20 - 25 °C 

25°C 

Coefficient for temperature compensation 

0…4 %/°C

 

tCC 

1.9 

TDS factor 

0.40 - 1.00 

Fct 

0.71 

Reset to factory default 

nO - YES 

rSt 

nO 

 

4.

 

Press 

 to enable the value changing of the selected parameter. 

5.

 

The value of selected parameter will start flashing. 

6.

 

Press   to change the value and then press 

 to confirm. 

7.

 

The value stops flashing. 

8.

 

Press   to select other parameters or press 

ESC

 to exit the setup menu. 

NOTE: Any time press 

ESC

 key to exit from SETUP menu. 

MEASUREMENT 

Rinse the electrode with distilled water or sample prior to start measurement. 
Fill the measuring cap with sample, switch on the meter with   and press 

MODE

 key to select desired 

parameter to be measured (green backlight during measurement). 
Immerse the Tester in sample and wait for stability, when stability icon 

 appears on display take the 

reading. 

During measurement make sure that pH electrode is free 

from air bubbles, and that there isn’t any air 

bubble around or between conductivity sensor. 

 

ORP  pH  SAL 
RES 

COND 

TDS 

-18.8.8 

µS 

mS 

pH 

mV 

18.8.8 

°C 

°F 

KΩ

 

MΩ

 

mg/L ppm ppt 

L  M  H 

HOLD 

pC 5 

r

 1.0 

COND 

tr

25 

°C 

COND 

2.0 

t

CC 

°C 

TDS 

0.7 1 

fct 

pH 

4.0 1 

pH 

23.5 

°C 

L  M  H 

ORP  pH 

SAL 

RES 

COND 

TDS 

-18.8.8 

µS 

mS 

pH 

mV 

18.8.8 

°C 
°F 

KΩ

 

MΩ

 

mg/L ppm ppt 

L  M  H 

HOLD 

Primary display 

 

Secondary display 

 

Summary of Contents for 1-5 Series

Page 1: ...Version 1 0 _ March 2016 Waterproof tester series Version 1 0 _ March 2016 INSTRUCTION MANUAL TESTER SERIES 1 5...

Page 2: ...WER OFF 5 SETUP MENU for pH1 Cond1 5 MEASUREMENT 5 CALIBRATION PROCEDURE FOR CONDUCTIVITY Cond1 6 CALIBRATION PROCEDURE FOR TDS Cond1 6 CALIBRATION PROCEDURE FOR pH pH1 6 INSTRUCTIONS FOR pH5 Cond5 PC...

Page 3: ...tion Accuracy Automatic scale 1 f s Automatic scale 1 f s MultiPoint Calibration 1 2 1 3 Temperature compensation 1 9 C 0 00 4 00 C Reference temperature 25 C 20 25 C TDS Range 0 1ppm 10ppt 0 1ppm 10p...

Page 4: ...ith fixed sensor and only two keys for all functions Tester series 5 is advanced version with replaceable sensor multicolour backlight display and 3 key for all functions SAFETY INSTRUCTIONS Read this...

Page 5: ...Cond TDS SAL During Setup Press to scroll in the menu or increase the value of the selected parameter CAL During measurement press to start the calibration of the selected parameter Press to confirm t...

Page 6: ...00 pH 1413 uS High 10 01 pH 12 88 mS First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with 4 x AAA Alkaline or NiMH rechargeable batteries Make sure the batterie...

Page 7: ...then the display will show TDS Fct flashing press to change this factor and then press to confirm Only for Cond1 5 The display will show rSt RESET nO flashing 6 Press and select YES if a reset of the...

Page 8: ...TDS with 1 or 2 points The calibration procedure for TDS is same as for Conductivity CALIBRATION PROCEDURE FOR pH pH1 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled wat...

Page 9: ...cted parameter 5 The value of selected parameter will start flashing 6 Press to change the value and then press to confirm 7 The value stops flashing 8 Press to select other parameters or press ESC to...

Page 10: ...pH5 PC5 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled water 3 Immerse the electrode in the 1st buffer solution pH7 00 and wait for stability 4 Press CAL key blue backl...

Page 11: ...istiller water or sample before use For short time storage use distiller water For long time store dry pH1 pH5 pH5 Food ORP5 PC5 Rinse the probe with distiller water or sample before use Store the pro...

Page 12: ...bbles in glass bulb 2 Check whether pH buffer solution is correct 3 Replace with new pH electrode Electrode slope out of range 85 or 110 pH measuring range out of range 0 00 pH or 14 00pH 1 Series 2 0...

Reviews: