Waterproof tester series 

5

 

INSTRUCTIONS FOR pH1 / Cond1

 

POWER ON 

Press   key once, the meter will switch On and the display will show all the segments active for 2 sec. and 
then it will display the following: 

 

 

 

  

  

All segments ON -> model name + software release -> measurement 
 

POWER OFF 

To turn off the meter press   key for 3 seconds, the meter will switch off. 

 

 SETUP MENU for pH1 / Cond1 

1.

 

With meter switched off, press and hold 

CAL

 key and press   key once. 

2.

 

The meter will switch on with all the segments active, release 

CAL

 key, the meter will go 

into the Setup Menu. 

3.

 

The display will show SLT on primary display with 

COND

 flashing (only for Cond1). 

4.

 

Press   to select the parameter between 

COND

 or 

TDS

 to be used for measurements and 

press 

 to confirm (Only for Cond1). 

o

 

Only if the 

TDS

 is selected then the display will show TDS Fct flashing, press   to 

change this factor and then press 

 to confirm (Only for Cond1). 

5.

 

The display will show rSt (RESET): 

nO

 flashing. 

6.

 

Press   and select 

YES

 if a reset of the meter is required and then press 

 to confirm. 

7.

 

At this point the meter finishes SETUP menu and switches off. 

NOTE: To skip the changing of the value simply confirm the flashing value with 

 key, the meter 

will go to the next Parameter. 
 
 
 

MEASUREMENT 

Rinse the electrode with distilled water or sample prior to start measurement. 
Fill the measuring cap with sample, switch on the meter with  , immerse the Tester in sample and wait for 
stability, when stability icon 

 appears on display take the reading. 

During measurement make sure that pH electrode membrane 

is free from air bubbles, and that there isn’t 

any air bubble around or between conductivity sensor. 

 

 

ORP  pH  SAL 
RES 

COND 

TDS 

-18.8.8 

µS 

mS 

pH 

mV 

18.8.8 

°C 

°F 

KΩ

 

MΩ

 

mg/L ppm ppt 

L  M  H 

HOLD 

pH  1 

r

 1.0 

pH 

4.0 

pH 

TDS 

0.7 1 

fct 

  No 

rst 

COND 

SLT 

Summary of Contents for 1-5 Series

Page 1: ...Version 1 0 _ March 2016 Waterproof tester series Version 1 0 _ March 2016 INSTRUCTION MANUAL TESTER SERIES 1 5...

Page 2: ...WER OFF 5 SETUP MENU for pH1 Cond1 5 MEASUREMENT 5 CALIBRATION PROCEDURE FOR CONDUCTIVITY Cond1 6 CALIBRATION PROCEDURE FOR TDS Cond1 6 CALIBRATION PROCEDURE FOR pH pH1 6 INSTRUCTIONS FOR pH5 Cond5 PC...

Page 3: ...tion Accuracy Automatic scale 1 f s Automatic scale 1 f s MultiPoint Calibration 1 2 1 3 Temperature compensation 1 9 C 0 00 4 00 C Reference temperature 25 C 20 25 C TDS Range 0 1ppm 10ppt 0 1ppm 10p...

Page 4: ...ith fixed sensor and only two keys for all functions Tester series 5 is advanced version with replaceable sensor multicolour backlight display and 3 key for all functions SAFETY INSTRUCTIONS Read this...

Page 5: ...Cond TDS SAL During Setup Press to scroll in the menu or increase the value of the selected parameter CAL During measurement press to start the calibration of the selected parameter Press to confirm t...

Page 6: ...00 pH 1413 uS High 10 01 pH 12 88 mS First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with 4 x AAA Alkaline or NiMH rechargeable batteries Make sure the batterie...

Page 7: ...then the display will show TDS Fct flashing press to change this factor and then press to confirm Only for Cond1 5 The display will show rSt RESET nO flashing 6 Press and select YES if a reset of the...

Page 8: ...TDS with 1 or 2 points The calibration procedure for TDS is same as for Conductivity CALIBRATION PROCEDURE FOR pH pH1 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled wat...

Page 9: ...cted parameter 5 The value of selected parameter will start flashing 6 Press to change the value and then press to confirm 7 The value stops flashing 8 Press to select other parameters or press ESC to...

Page 10: ...pH5 PC5 1 Power ON the meter by pressing the Key 2 Rinse the electrode with distilled water 3 Immerse the electrode in the 1st buffer solution pH7 00 and wait for stability 4 Press CAL key blue backl...

Page 11: ...istiller water or sample before use For short time storage use distiller water For long time store dry pH1 pH5 pH5 Food ORP5 PC5 Rinse the probe with distiller water or sample before use Store the pro...

Page 12: ...bbles in glass bulb 2 Check whether pH buffer solution is correct 3 Replace with new pH electrode Electrode slope out of range 85 or 110 pH measuring range out of range 0 00 pH or 14 00pH 1 Series 2 0...

Reviews: