M U L T I - A N G L E S P E C T R O P H O T O M E T E R
47
Instrument Specifications
General
Measurement Geometries
Illumination:
45°
Aspecular Viewing:
-15° (MA96, MA98), 15°, 25°, 45°, 75°, 110°
Out-of-plane (MA98):
25°az90°, 25°az-90°, 60°az125.3°, 60°az-125.3°
Secondary Illumination (MA98): 15°
Aspecular Viewing (MA98):
-15°, 15°
Angular Accuracy:
±0.15° Fiber Optic pick up coupled with DRS technology
Measuring Area:
Approx. 12 mm (.5")
Light Source:
Gas filled tungsten lamp
Lamp Life:
750,000 measurements, typical
Spectral Range:
400nm – 700nm
Spectral Interval:
10nm (31 measured points)
Measurement Range:
0 - 400%
Colorimetric Illuminants:
A, C, D50, D65, F2, F7, F11 and F12
Colorimetric Standard Observers: 2° & 10°
Colorimetric Scales:
L*a*b*, L*C*h°,
∆
E*,
∆
E
CMC
,
∆
E DIN6175,
∆
E
2000
Effect Parameters:
xDNA, Flop Index
Measurement Time:
Approx. 2 seconds
Reproducibility:
0.18 ΔE* avg on reference Series II BCRA tile set
(Inter-instrument agreement)
Repeatability:
0.03
∆E*
on white cal plaque (20 measurements at 5 sec
intervals)
Power Supply:
Rechargeable Lithium Ion battery pack 7.4vDC @
2400 mAh
AC Adapter:
Input 100-240V 50-60 Hz
Output 12VDC @ 2.5 A
Measurements per charge:
Up to 750
Measurement storage:
250 Standards
1000 Samples
Data interface:
USB 2.0
Weight:
2.5 lbs (1.13 kgs)
Dimensions:
3.4 x 4.5 x 10.6 (8,7 cm x 11,4 cm x 26,9 cm)