M U L T I - A N G L E S P E C T R O P H O T O M E T E R
26
3.
Use the
Right
navigation button to move the highlight focus to the Progress column. Use the
Down
navigation button to highlight the Next arrow icon (
→
) if not highlighted and press the
Select
button.
4.
Remove the hole plug and snap the instrument on the black trap as previously explained.
Make sure the Next arrow icon (
→
) is highlighted in the Progress column and press the
Select
button.
The instrument takes 4 or 8 measurements depending on the model. After the black
calibration measurements are complete, remove the instrument from the black trap, install the
hole plug, and return the black trap to its storage location.
NOTE:
If an error message appears during or after black trap calibration, press the
Select
button to clear the message and try measuring the black trap again. If an error still occurs,
clean the black trap as explained in the Appendices.
5.
Position the instrument on the white reference as previously explained and press the
Select
button. Take the measurement(s) by applying the appropriate amount of pressure to the port
to activate the pressure sensors (green indicator LEDs). Hold the instrument steady
throughout the measurement sequence.
The instrument takes 1 or 2 measurements depending on the model. After the white
calibration measurement(s) is complete, press the
Select
button to return the display to the
main screen.
Remove the instrument from the calibration reference and return the reference to its storage
location.
NOTE:
If an error message appears during or after white calibration, Press the
Select
button
to clear the message and try measuring the white reference again. If an error still occurs,
clean the white calibration reference as explained in the Appendices.