M A 3 S P E C T R O P H O T O M E T E R
22
Contact sensor(s) are not
working correctly. Sensors
may be stuck due to dirt or
paint; or they are defective.
Open a new job to show the 3 sensor indicators
on the screen and place the instrument on a flat
surface and then lift it. If one or more of the
sensor indicators in the screen do not change
color when lifted, there may be a problem with
the contact sensor(s). Contact technical
support.
Shutter error.
Instrument defective.
Contact technical support.
Multiple error messages.
Instrument encountered a
cascade of errors.
Press OK button in error message for 2
seconds.
Date and time wrong.
Incorrect setting.
Refer to instruction in manual.
Instrument will not
measure or calibrate.
Sensor indicators in the
display do not change color.
Contact sensors are not working correctly. Put
the instrument into measure mode and place
the instrument on a flat surface and then lift. If
one or more of the sensor
indicator
s in the
display do not change color when lifted, there
may be a problem with the contact sensor.
Contact technical support.
Shutter error.
Instrument defective.
Contact technical support.
Multiple error messages.
Instrument encountered a
cascade of errors.
Press OK button in error message for 2
seconds.
Date and time wrong.
Incorrect setting.
Refer to instruction in manual.
Instrument Specifications
Environmental
Operating Temp:
50°F to 95°F (10°C to 35°C)
Humidity Max:
85% RH max (non condensing)
Storage Temp:
-4°F to 122°F (-20°C to 50°C)
Battery
Type:
Lithium Ion Battery
Cell Size:
18650
Nominal Voltage:
3.6V
Capacity:
2.15 Ah
Max. discharge current:
4A
Max. charge current:
2.15A
Inner resistance:
<= 120 mΩ
Compliance:
UN38.3, IEC62133B, CE
n:
Protection:
over current, over charge, over discharge