EasyTrax Setup and Operation
12
Laser Alignment
CAUTION:
CLASS 1 LASER PRODUCT
1.
Manually align the color bar following the Manual Color Bar Alignment procedure discussed
earlier.
2.
Position the read head reticule over the first patch on the color bar as previously explained.
3.
Launch the Press Tool application and turn on the laser line by pressing the
Laser
button on
the read head (the laser automatically turns off after 30 seconds).
•
If the laser line is not centered down the length of the color bar, continue with step 4.
•
If the laser line is centered down the length of the color bar, press the Laser button
again to turn it off. The alignment procedure is not required
. You are now ready to
perform color bar measurements.
4.
Note: Only continue with these steps if an alignment is required.
If the laser was not aligned, remove the supplied 1.5 mm hex key from its storage
compartment at the back of the vertical arm.
Laser button
Laser line along color bar
1.5 mm hex key
storage location