Press the instrument firmly to the shoe and hold steady until the display indicates that the
reading is complete. Release the instrument.
10.
Tap the Select control to save the calibration and exit to the main menu.
11.
Store the calibration reference in a dry, dust free area, away from direct exposure to light.
NOTE: If an error message appears during or after white calibration, tap the Select control to
clear the message and try measuring the white reference again. If an error still occurs, clean
the white calibration reference as explained in the Appendices.