4
2
Calibration and Measurement
FREQUENCY OF CALIBRATION
Typically, the host computer prompts for an instrument calibration when required. The frequency at which this occurs
depends on the application. If desired, a reflection calibration can be invoked manually by pressing the Instrument
button—refer below for procedure. At a minimum the instrument should be calibrated once a day for best accuracy.
REFLECTION MODE CALIBRATION
The X-RiteColor AutoScan Spectrophotometer has a unique automatic calibration feature. When the instrument performs a
calibration procedure, the Color Reflection Reference is automatically scanned, simplifying the procedure. The instrument
employs a number of self-checking procedures to verify calibration accuracy during normal use. When internal limits are
exceeded, the computer calls for a calibration.
CALIBRATION NOTE:
Handle Color Reflection Reference by the edges. Make sure
that the Color Reflection Reference is free of dust, dirt, and
smudgemarks. Refer to Section Three for cleaning procedure.
Always store the Color Reflection Reference in its protective
envelope away from light and heat.
1.
Center the designated end of the Color Reflection
Reference under the alignment mark. Slowly insert strip
past the front idler rollers up to the black bar on the
reference strip.
2.
To manually force the instrument into calibration mode,
press and hold the Instrument button for a minimum of three
seconds.
The indicator light changes to yellow. Note, if calibration is
not performed after a short period of time, the instrument returns to
the measurement mode (green light).
3.
Press the Instrument button again. The light slowly flashes yellow.
After a few moments, the light changes to green and then back to
slow flashing yellow. The strip is then pulled through the instrument
and out the back. The light turns solid green after a successful
calibration. If the calibration fails (fast flashing yellow light), verify
strip is clean and re-read.
4.
Place Color Reflection Reference in its protective envelope.
NOTE:
The optics block must remain in the T/W position when
calibrating a DTP41/T instrument.
TRANSMISSION MODE (BLACK) CALIBRATION
NOTE: Transmission operation is optional. Therefore, transmission mode calibration may not be available on your
instrument.
A transmission calibration is recommended for the DTP41/T instrument approximately once a week for optimal
performance. The host computer initiates the procedure and prompts you to insert the calibration envelope (transmission
calibration cannot be initiated manually). This procedure guides you through the mechanical aspects of positioning the
envelope in the instrument.
1.
When the host computer indicates a transmission calibration is needed, center the designated end of the Color
Reflection Reference Envelope under the alignment mark. Slowly insert past the front idler rollers until it comes to
rest against the rear drive rollers (approx. three inches).
2.
Activate the cal procedure at the computer
. NOTE: This is a static measurement, the envelope is not pulled through
the instrument.
•
The instrument beeps once, flashes yellow slowly and takes a series of five measurements. After completion, the
instrument beeps again and the indicator light changes to solid green.
3.
Remove the Color Reflection Reference Envelope from the Instrument.
Press button
down for 3 sec. to
initiate calibration,
then press button
again
Slowly insert strip up
to the black bar
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for AutoScan DTP41 Series
Page 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 16: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...