clino_e.doc
Page 8 of 24 Pages
– 7 –
Battery indicator
Low battery voltage (drained battery) is indicated by the sign "BATT" appearing in inverted mode
at the lower left corner of the display.
- 8 -
Interface port RS 485
True RS485 – Interface / connections to various WYLER instruments such as e.g.
Levelmeter 2000, WYLER T/C as well as to a RS 232 com port of a PC (Special WYLER
connector required)
2.3 B
ATTERY TEST
/
C
HANGE OF
B
ATTERY
Appearance of the sign "BATT" in the lower left corner of the display, this will be displayed constantly in all
measuring modes, indicates a drained battery, which must be replaced.
The display "BATT" allows for timely replacement of the battery.
Specification:
•
1 x Size AA 1.5V Alkaline / 130 hrs
•
1 x Size AA 3.0V MnO2-Lithium / 300 hrs
Battery installation
2.4 F
UNCTION CHECK
Press the key
"ON/MODE" until all the display's segments are visible. Immediately after releasing the key the
display shows
until the first measuring value is seen in the last selected measuring mode.
Incline the +CLINO PLUS+ to the left. The following displays are correct:
(any value)
Incline the +CLINO PLUS+ to the right. The following displays are correct:
(any value)
The +CLINO PLUS+ persistently indicates:
flashing means that no calibration data is available. Data
was probably lost or destroyed by accidentally calling the calibration utility. The +CLINO PLUS+ must be
recalibrated. See chapter 5.1 for calibration procedure. For possible salvation see chapter 2.10.
2.5 S
ET ABSOLUTE ZERO
The absolute zero represents a base for absolute inclination measurements (Deviation from horizontal or vertical).
In order to achieve best results take care that the +CLINO PLUS+ and measuring object are at identical
temperatures and put the +CLINO PLUS+ into operation a few minutes prior to zero setting.
The absolute zero is automatically calculated and set from the two values entered whilst conducting a reversal
measurement. (Reversal measurement = Two measurements made in the same spot but in exactly opposing
directions). For this operation place the +CLINO PLUS+ upon a suitable surface, (rigid location; as flat as possible;
as near to horizontal as possible). In order to allow, positioning in exactly the same location, after rotating the
instrument through 180 degrees, mark out position and particularly the orientation of the +CLINO PLUS+.