2.2 Environmental conditions
Operation is only permitted indoors . The device must be protected from moisture, rain and splash water .
Permissible environmental conditions for operation:
Ambient temperature
+5°C to +35°C (41°F to 95°F)
Relative humidity
0–85% (not condensing)
Altitude
Max . 2000 m over normal zero
Degree of pollution
2
Permissible environmental conditions for transport:
Ambient temperature
–10 bis 45°C (14°F to 113°F)
Relative humidity
0–85% (not condensing)
2.3 Supply Voltage
Only power packs approved by the manufacturer may be used to supply the device . In particular, it must be
ensured that the following specifications are adhered to.
Input voltage
85 ∼ 264 V/AC, 47–63 Hz
Output voltage
24 V/DC (+/– 3%)
Output current/
output power
3 A/72 W
Power adapter
Only the power supply unit provided by the
system supplier may be used .
Input connector
IEC socket IEC60320-C14
connector 24 V output
2 .1 mm plug (, outside –)
2.4 Main Power cable
Depending on the place of use, the power supply unit can be connected to the respective regional power
supply plug with different connection cables . There is an IEC60320-C14 socket (3-pin IEC connector) on
the power supply unit itself .
Only cables that comply with IEC60320 may be used .
Minimum cross-section in mm²
0 .75 mm²
Minimum cross-section in AWG
18 AWG
Maximum length
3 m