Operation
pH 3300i/3400i
42
ba75682e02
03/2008
4.5
Transmitting data (only pH 3400i)
You have the following possibilities of transmitting data:
z
One of the following options:
– With the
AutoStore
function (page 34), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the
Data transmission interval
function (Int 2), measured values
are periodically output on the interface (see below).
z
With the
Output data storage
function (page 36), calibration data or saved
measured values are output on the interface.
z
Via the analog recorder output (page 44), measured values are output as
voltage values.
z
With the KOM pilot communication kit (accessory), data can be transmit-
ted bidirectionally (page 45).
Note
If you connect a recorder (analog output), the output on the digital interface
is switched off.
4.5.1
Data transmission interval (Int 2, pH 3400i)
The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface.
Note
When the
AutoStore
function is active, the data transmission is performed
according to the setting of the save interval (Int 1). Set the save interval (Int 1)
to OFF to activate the
Data transmission
interval (Int 2).