Operation
pH 3300i/3400i
46
ba75682e02
03/2008
4.6
Configuration
You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):
Note
You can leave the configuration menu at any time with
<M>
. The parameters
that have already been changed are stored.
Baud rate
(pH 3400i)
Baud rate
1200, 2400,
4800
, 9600
Display during the
pH calibration
Buffer nominal value
, or current
electrode voltage
Calibration interval (Int 3)
1 ...
7
... 999 d
Date/time
Any
pH resolution
0.01
or 0.001
1
Switch off the measuring instrument.
2
Press the
<M>
key and hold it down.
3
Press the
<ON/OFF>
key.
The display test appears briefly on the display. The measuring instru-
ment then switches automatically to the setting of the baud rate.
4
Select the required Baud rate with
<
▲
> <
▼
>
.
5
Confirm with
<RUN/ENTER>
.
CAL diSP
appears on the display.
Day.Month No.
Time
1
2
M
m
%
mg/l
cm
S/
cm
S
mbar
mV/pH
t
n
Store
Auto
LoBat
Tref25
nLF
/
K
%
RCL
Sal
°
AR
TP
C
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
O
UpH
AutoCal DIN
AutoCal TEC
Ident
Year
0
0
8
4
Baud