Operation
pH 3300i/3400i
34
ba75682e02
03/2008
4.4.2
Saving automatically
The save interval (Int 1) determines the chronological interval between auto-
matic save processes.
After the fixed interval has expired, the current data record is transmitted to
the storage and to the interface.
Setting the save
interval
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min):
1
Press the
<RUN/ENTER>
key and hold it down.
2
Press the
<STO>
key.
Int 1
appears on the display.
3
Set the required interval between the saving procedures with
<
▲
>
<
▼
>
.
4
Confirm with
<RUN/ENTER>
.
The number of free memory locations appears on the display.