pH 3300i/3400i
Operation
45
ba75682e02
03/2008
4.5.3
PC/external printer (RS 232 interface, pH 3400i)
Via the RS 232 interface, you can transmit the data to a PC or an external
printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the inter-
face to the instruments.
The data output automatically switches to the
RS 232
interface.
Warning
The RS 232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Socket
assignment
4.5.4
Remote control (pH 3400i)
The measuring instrument can be remotely controlled from a PC. This re-
quires the KOM pilot communication kit. It is available as an accessory.
The instrument is then controlled via commands that simulate keystrokes and
request the current display contents.
Note
A more detailed description is provided within the scope of delivery of the
communication kit.
Baud rate
selectable between: 1200, 2400,
4800
, 9600
Handshake
RTS/CTS + Xon/Xoff
Parity
none
Data bits
8
Stop bits
1
1 CTS
2 RxD
3 Ground
4 TxD
1
2
3
4
RS 232
REC