What to do if...
Multi 350i
86
6.3
DO measurement
DO sensor was not
recognized
Error message
OFL
Error message,
Error
(invalid calibration)
Error message,
LEAK
Measured value
flashes
for a long time
(No stable measured
value)
Measured value too low
Cause
Remedy
– Sensor not connected
– Connect sensor
– Cable broken
– Replace sensor
Cause
Remedy
– Measuring range exceeded
– not possible
Cause
Remedy
DO sensor:
– Electrolyte solution depleted
– Regenerate sensor
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate sensor
Cause
Remedy
DO sensor:
– Membrane damaged
– Exchange the membrane cap
– Membrane head screwed on
not tight enough
– Screw membrane head tight
Cause
Remedy
DO sensor:
– Membrane contaminated
– Clean membrane
Cause
Remedy
DO sensor:
– Insufficient flow
– Provide flow to the sensor