Multi 350i
Operation
65
5
Press
<MENU/ENTER>
.
The AutoRead measurement to determine the cell constant
starts.
6
Wait for the end of the AutoRead measurement or accept the
calibration value with
<MENU/ENTER>
.
The determined cell constant is displayed as a message (
i
).
7
You have the following options:
!
Accept the new calibration values with
<MENU/ENTER>
.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
!
To switch to the measured value display without accepting
the new calibration values, press
<M>
or
<ESC>
.
i
Cell constant = 0.481 1/cm
i
Temperature = 24.8 °C
Terminate AutoRead
<CAL>
i
Cell constant = 0.481 1/cm
i
++
Accept
<CAL>