![wtw IS 602 Instruction Manual Download Page 20](http://html1.mh-extra.com/html/wtw/is-602/is-602_instruction-manual_1000604020.webp)
Measurement
Interferences / Accessories
20
BA31110/07.00/AS/IS 602
¾
1
BA31110/07.00/AS/IS 602
¾
1
21
•
Measure sample volume (fill the measuring
beaker completely) and pour into one of
the light-proof sample bottles.
•
Put a magnetic stirring rod into each filled
sample bottle.
•
Insert a rubber quiver into each bottle neck
and put 1 or 2 sodium hydroxide tablets
into the quiver with tweezers.
•
Put on closing cap but do not yet screw
tight.
•
Connect line adaptor to mains and to the
instrument.
•
Close incubator; for temperature
adjustment (20°C) stirr for approx. 1 hour.
•
Close the caps of the sample bottles and
storage beakers airtight.
•
Align zero point of the scale with meniscus
of the mercury column.
•
Note date and time in the data form sheet.
•
Every day read the values and note them
in the diagram note block. At the end of the
measurement (usually after 5 days) read
scale value and multiply with the
conversion factor.The result corresponds
to the oxygen consumption in mg/l.
•
After measurement screw the closing caps off the sample bottles and loosen
the caps of the mercury storage beakers.
•
Rinse rubber quiver, sample bottles and magnetic stirring rods only with warm
water, before measurement rinse them with sample.
Note
Add reagents for sample preparation to the sample before measuring
the sample volume. With volume additions of more than 5 ml per liter
take the created dilutions into account for your result.
Close the mercury storage beaker with the bottle cap if it is not used
for a longer time.
Interferences
•
The scale shows zero or too low a value.
The measuring equipment is not airtight.
Check rubber quiver, screw top and hose connection.
•
Measured value below zero.
The temperature of the sample had not been sufficiently adjusted (< 20°C).
•
Measuring range exceeded.
The measuring range chosen was too small. With very high values > 4000 mg/l
dilute the sample.
Errors due to procedure have not been mentioned.
Recommended accessories
Model
Order no.
Marking ring set (6 pcs.) with label set (1-12) for
bottles
Mark-6
209 013
Overflow measuring beaker
with volume
22.7 ml
43.5 ml
97.0 ml
164 ml
250 ml
365 ml
432 ml
MK 22/60
MK 43/600
MK 97/600
MK 164/600
MK 250/600
MK 365/600
MK 432/600
209 030
209 040
209 050
209 060
209 070
209 080
209 090
Spare parts / Auxiliary material
Model
Order no.
Sample bottle brown
PF 600
209 100
Stirring rods
RST 600
209 120
Rubber quiver
GK 600
209 170
Stirring rod remover
REF 600
209 130
Sodium hydroxide tablets
NHP 600
209 140
Nitrification restraint
(allylthiourea solution 250 ml)
ATH 600
209 330
Diagram note blocks
Dia/600
209 020
Summary of Contents for IS 602
Page 11: ...Technische Daten BA31110 07 00 AS IS 602 1 11 Pr fzertifikat EMV IS 6...
Page 12: ...Technische Daten 12 BA31110 07 00 AS IS 602 1 Pr fzertifikat EMV IS 12...
Page 13: ......
Page 23: ...Technical data BA31110 07 00 AS IS 602 1 23 Test certificate EMC IS 6...
Page 24: ...Technical data 24 BA31110 07 00 AS IS 602 1 Test certificate EMC IS 12...