Contents
Instruction manual
Manometric BOD Measuring Instrument IS 602
Putting into operation ................................................................................................17
Filling ...................................................................................................................17
Mains supplied operation ....................................................................................18
Operation notes .........................................................................................................19
Notes .........................................................................................................................19
Measurement.............................................................................................................19
Interferences .............................................................................................................21
Accessories ...............................................................................................................21
Recommended accessories ................................................................................21
Spare parts / Auxiliary material ............................................................................21
Technical data ...........................................................................................................22
Test certificates EMC...........................................................................................23
Filling
When filling the mercury storage beakers work very carefully!
Mercury is extremely poisonous and easily volatile. For safety
reasons perform the filling in a big plastic tub.
Collect spilled mercury completely with tin foil, a brush or a mercury
collecting pipette.
If other methods do not work for collecting small amounts of mercury
which have possibly penetrated into gaps: Cover them with a layer of
mercury binding agent several millimeters high. This will prevent
volatilization.
Store contaminated mercury remnants in a suitable and closed bottle
and take them to a disposal center.
Ÿ
Incline the storage beakers and with the
funnel carefully pour a 2 ml mercury
portion in each of the mercury storage
beakers .
The filling of the glass capillary should be
free of bubbles.
Ÿ
Put on closing caps, but do not yet tighten
them.
Ÿ
Check for air bubbles:
With a Peleus ball which has been put on
the hose connector (connection to glass
capillary) take in mercury approx. to the
end of the scale.
Ÿ
Remove air bubbles by pressing the
mercury back into the storage beaker (with
Peleus ball).
Ÿ
Pour 1 ml deionized water onto the
mercury in the storage beaker.
From time to time replenish the water.
Ÿ
With the screw cap close the storage
beaker airtight.
Summary of Contents for IS 602
Page 11: ...Technische Daten BA31110 07 00 AS IS 602 1 11 Pr fzertifikat EMV IS 6...
Page 12: ...Technische Daten 12 BA31110 07 00 AS IS 602 1 Pr fzertifikat EMV IS 12...
Page 13: ......
Page 23: ...Technical data BA31110 07 00 AS IS 602 1 23 Test certificate EMC IS 6...
Page 24: ...Technical data 24 BA31110 07 00 AS IS 602 1 Test certificate EMC IS 12...