MIQ/CR3; DIQ/CR3
Settings
4 - 41
ba64107e08
05/2006
4.9.1
Switching on the maintenance condition
4.9.2
Switching off the maintenance condition
1
Call up the measured value display with
m
.
2
Select the sensor you want to switch on the maintenance
condition for with
d
. The measured values of the sensor do
not flash.
3
Open the
Display/Options
menu with
g
.
4
Using
d
, highlight the menu item,
Switch maintenance
condition on/off
or
Maintenance of sensor S0x
and confirm with
g
. A window that informs you about the maintenance
condition opens up.
5
Confirm
Continue
with
g
.
The selected sensor is in the maintenance condition. Linked
outputs are frozen.
6
Call up the measured value display with
m
.
The measured values of the sensor flash.
1
Call up the measured value display with
m
.
2
Select the sensor you want to switch off the maintenance
condition for with
d
. The measured values of the sensor
flash.
3
Open the
Display/Options
menu with
g
.
4
Using
d
, highlight the menu item,
Switch maintenance
condition on/off
or
Maintenance of sensor S0x
and confirm with
g
. A window that informs you about the maintenance
condition opens up.
5
Confirm
Continue
with
g
.
The maintenance condition of the selected sensor is switched
off. Linked outputs are released.
6
Call up the measured value display with
m
.
The measured values of the sensor do not flash.