AmmoLyt
®
700 IQ
3 - 8
ba75353e05
03/2006
Carrying out settings
Temp. adjustment
-1.5 °C ... +1.5 °C
The temperature compensation function
enables the temperature sensor to be ba-
lanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
!
Due to the thermal capacity of the sen-
sor, it is necessary to place it in a con-
tainer with at least 2 liters of water.
!
Leave the sensor in this container for at
least 15 minutes while stirring occasio-
nally, then carry out the adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.
Potassium compens.
On
Off
Potassium ions in the test sample interfere
with the measurement and lead to increa-
sed values (see section 4.2.2).
After determining the potassium content of
the test sample by a reference measure-
ment you can input the determined potas-
sium content here.
The potassium compensation corrects the
measured value accordingly.
Potassium conc.
(only with
Potassium
compens.
:
On
)
0 ... 100 mg/l
Save and quit
The system confirms the saving of the set-
tings and the display switches to the next
higher level.
Quit
The display switches to the next higher
level without saving the new settings.
Menu item
Possible settings
Explanations
1
Switch to the measured value display with
m
.
2
Open the
Settings
menu with
s
.
3
Select and confirm the menu item
Settings of sensors and diff.
sensors
->
Measuring range
column with
d
and
g
.
4
Select an entry with
d
.
Summary of Contents for AmmoLyt 700 IQ
Page 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Page 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Page 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Page 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Page 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...