Calibration and measuring
AmmoLyt
®
700 IQ
4 - 20
ba75353e05
03/2006
Calibration history
Fig. 4-6
330 - Calibration history of selected sensor
The calibration history contains the following information:
Note
Only calibration data produced by the same calibration procedure can
be compared.
Calibration data of
the initial calibra-
tion.
List with calibration
data of the last cali-
brations
Date
Date of the calibration
S(*)
Slope [mV] of the electrode
The value is marked by
* Values unchanged
if it is not possible to
determine the slope. Instead, the last determined value is taken over
for the slope.
This applies to all 1 point calibration procedures,
1 point standard (1)
,
1 point ref. (2)
and
Simple std. add. (4)
.
DV
Drift potential [mV]
Ref1
/
Ref2
Concentration [mg/l NH4-N]
!
of the standard with
1 point standard (1)
!
of the test sample (reference measurement) with
1 point ref. (2)
!
of the two standards with
2 point stand. (3)
!
of the test sample (calculated) with
Simple std. add. (4)
[Ref1/-]
!
of the test sample (calculated) with
Double std. add. (5)
[Ref1/
Ref2]
K+
Potassium content [mg/l]
P
Calibration procedure, number 1 ... 5
T
Temperature [°C]
R
Result of the calibration
+
: Calibration successful
-
: Calibration unsuccessful
Summary of Contents for AmmoLyt 700 IQ
Page 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Page 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Page 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Page 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Page 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...