![WPI UltraMicroPump II Instruction Manual Download Page 13](http://html.mh-extra.com/html/wpi/ultramicropump-ii/ultramicropump-ii_instruction-manual_999839013.webp)
UltraMicroPump II
World Precision instruments
11
Type
Syringe Volume
(defined for this length)
ID (mm)
nL / step REV J
Max. Rate
nl /sec or
nl /min
A
0.5 µl (54.1 mm)
†
0.1085 (0.102 reV H)
0.0294 ( 0.026 rev H)
20
B
1.0 µl (54.1 mm)
†
0.1534 (0.145 reVH)
0.0587 (0.052rev H)
40
C**
5 µl (54.1 mm)
†
0.343 (0.325 reV H)
0.2934 (0.263 rev H)
202
D
10 µl (54.1 mm)
†
0.485 (0.46 reV H)
0.5868 (0.528 rev H)
406
E
25 µl (60.0 mm)
0.73
1.329
1022
F
50 µl (60.0 mm)
1.03
2.646
2035
G
100 µl (60.0 mm)
1.46
5.315
4088
H
250 µl* (60.0 mm)
2.3
13.191
9999
I
500 µl* (60.0 mm)
3.26
26.501
9999
J
1000 µl* (60.0 mm)
4.61
52.995
9999
K
nanoliter 2000
† †
0.48 plunger in 0.50 glass
2.3 nl /step (0.0005” step)
884
L
10 µl (60 mm)
0.4607 (0.100 rev H)
0.5293 (0.0249 rev H)
170
M
user defined
see page 12
**
ils005 (27 mm)
0.4856
0.5880 compensates for length as
tYPe m
* Gas-tight syringes are not recommended for UMPII in these volumes; instead, use a liquid-tight
syringe to prevent drive motor damage.
** The ILS005 5 µL syringe must use type M with a 0.5880 nL / step entry. See page 15 for details.
†
These are defined as 60 mm in Rev. H firmware.
† †
WPI’s
Nanoliter 2000
, a nanoliter injector for the 2-70 nL range, comes with its own simple controller
but may also be driven by the Micro4. For more information, enquire about WPI #
B203XVY
.
Note On Syringe Stroke Length
the delivery of the umPii is based on 60 mm or 54.1 mm syringes. Please note
which syringe length you are using as a factor of 0.9016 may need to be applied to
the volume to be injected in or-
der to have a precise injection.
For the ils005, use the tYPe
“m” values in the above table.
Maker
Syringe
Stroke
Length
Use TYPE
Hamilton
1700 series ,10 µl
60 mm
L
Hamilton
700 series,10 µl
54.1 mm
D
sGe
0.5 µl – 10 µl
54.1 mm
A,B,C,D
ils 5 µl luer tip
ils005
28 mm
M
sGe
25 µl – 500 µl
60 mm
E - L
WPi
FlexiFil™
54.1 mm
D
WPi
nanoFil™
60 mm
L
*Manufacturers have the right to
change this specification at any time.