STX2-PLUS

World Precision Instruments 

1

ABOUT THIS MANUAL

The following symbols are used in this guide:

This symbol indicates a CAUTION. Cautions warn against actions that can cause 

damage to equipment. Please read these carefully.

This symbol indicates a WARNING. Warnings alert you to actions that can cause 

personal injury or pose a physical threat. Please read these carefully.

NOTES and TIPS contain helpful information.

Fig. 1—The STX2-PLUS electrode is designed for use with the EVOM3 meter.

INTRODUCTION

The

 STX2-PLUS

 electrode can be used with the 

EVOM3

 and is designed to facilitate 

measurements of membrane voltage and resistance (TEER) of cultured epithelia 

directly in 12- and 24-well tissue culture plates. The electrode incorporates a 

adjustable pair of probes, 5 mm tall, 4 mm wide and 0.7 mm in thickness. Each probe 

has an outer (current) and an inner (voltage) electrode. 
For 12- or 24-well measurements, the 

STX2-PLUS

 electrode is capable of making 

precise measurements. The 

STX2-PLUS

 is designed primarily for 24-well inserts, and 

it can be used with 24-well HTS plates (Corning 3378) and 12-well inserts (Corning 

3460) with good results, as long as the placement is consistent. Consistent electrode 

placement and mechanical stability are essential for consistent resistance results. Try 

to keep the fluid levels and fluid temperatures constant. The use of a warming plate 

can add to measurement items and reduce CO

2

 out gassing during readings. CO

2

 out 

gassing can change the pH of the media and sharply increase a resistance reading.

Summary of Contents for STX2-PLUS

Page 1: ...WORLD PRECISION INSTRUMENTS STX2 PLUS for use with EVOM3 Epithelial Voltohmmeter Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 082620...

Page 2: ......

Page 3: ...NG INSTRUCTIONS 3 Positioning Electrode in a Culture Cup 3 Stable Repeatable Readings 3 Ease of Use 3 Adjusting the Length of the Electrodes 4 MAINTENANCE 5 Cleaning Maintaining the STX2 PLUS Electrod...

Page 4: ...ii World Precision Instruments...

Page 5: ...rectly in 12 and 24 well tissue culture plates The electrode incorporates a adjustable pair of probes 5 mm tall 4 mm wide and 0 7 mm in thickness Each probe has an outer current and an inner voltage e...

Page 6: ...any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Department...

Page 7: ...ility of the measurements is significantly improved Fig 2 Right Electrodes in solution Stable Repeatable Readings This new electrode design sets the correct depth and centering of the electrode elemen...

Page 8: ...a hanging insert Fig 6 Rotate the nut to adjust the length of the electrodes Fig 7 The electrode length is properly adjusted 2 Align the electrodes with the well inserts and position the STX2 PLUS ove...

Page 9: ...to your maintenance schedule and the frequency of your cleaning Soak the electrodes overnight when the electrodes have not been on a routine maintenance cleaning schedule Soak 60 minutes if you are o...

Page 10: ...us electrode must be dipped into chlorine bleach 5 sodium hypochlorite on occasion to maintain the amount of Ag to AgCl balance Otherwise the electrode will become unstable A 10 15 minute immersion of...

Page 11: ...should show an R2 value 0 99x NOTE These resistance readings can vary due to fluid volume and electrode depth These values are a reference An Excel scatter plot of the molarity vs resistance as a pow...

Page 12: ...ng Radio frequency interference Turn off or move any cellular phones farther away from the experimental setup NOTE If you have a problem issue with that falls outside the definitions of this troublesh...

Page 13: ...25 mL of 20 mM KCl 25 mL 50 mL 10 mM KCl APPENDIX B COMPATIBLE INSERTS AND PLATES Corning Millipore Material Membrane Diameter mm Growth Surface Area cm2 Membrane Pore Size m 3470 6 5 0 33 0 4 3472 PI...

Page 14: ...10 World Precision Instruments Millicell Membrane Pore Size m Qty Pkg MCHT24H48 0 4 48 MCRP24H48 1 0 48 MCSP24H48 3 0 48 MCMP24H48 5 0 48 MCEP24H49 8 0 48...

Page 15: ...special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or i...

Page 16: ...Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong New District Shanghai PR China...

Reviews: