Removing Water from the Surgical Bed
Water in the fluid system should be changed monthly or more often depending upon
use.
1. Hold the surgical bed in a vertical position with the fittings facing upward.
2. Disconnect the tubes from the pump.
3. Hold the tubes in a vertical position with the fittings facing upward.
4. Hold the tubes over a sink and remove the fittings from the tubes, allowing the
water to drain into the sink.
5. Turn the surgical bed vertically with the fittings end down, allowing the water to
flow from the surgical bed into the sink.
Removing Water from the Water Pump
1. Be sure the tubes are disconnected from the pump.
2. Unscrew the cap on the top of the water pump and tip the pump over a sink until
all the water is emptied from the pump.
3. Replace the cap on the pump.
Disinfecting the Water Pump and Surgical Bed
The water pump and surgical bed fluid system should be disinfected monthly, using
the following procedure:
1. Drain the water from the system following previous instructions.
2. Add a quaternary disinfectant solution to the pump reservoir, following the
instructions for filling the system from “Water Pump And Surgical Pump Setup” on
page 5.
3. Set the pump temperature to its lowest setting and run the system for one hour.
4. Drain the system of disinfectant solution and refill it with distilled water.
Cleaning the Surface of the Pump
Choose one of the following methods for cleaning the surface of the pump:
•
Use a damp cloth with soapy water.
•
Use Fantastik
®
spray cleaner.
•
Use a mild, non-bleach abrasive, such as Soft Scrub
®
.