background image

World Precision Instruments

2.  Set the vaporizer concentration to 5% until the animals become immobile. 
3.  Then, reduce the vaporizer setting to the species-appropriate level. Set the 

Isoflurane vaporizer dial to:

• 

Mice: 1.5%

• 

Rats <300gms: 2%

• 

Rats >300gms: 2 – 3%

• 

Guinea Pigs: 5% 

NOTE

: Guinea pigs should be anesthetized at 5%. Reduce the vaporizer setting to 3% 

before transferring the animal to the surgical bed.

NOTE

: When performing surgery longer than one hour, it may be necessary to reduce 

the Isoflurane vaporizer dial setting by 0.5-1.0% to prevent the animals from being 

exposed to too much gas.
4.  To be sure of complete anesthesia, animals should remain in the induction 

chamber for 3–5 minutes after initial induction before moving them to the 

surgical breathing unit. Animals in the induction chamber will remain anesthetized 

for as long as the gas flow continues.

MAINTENANCE

1.  Once animals are ready to be moved to the breathing unit, be sure that the 

desired breathing units are connected to the manifold.

2.  Reduce the secondary flowmeter setting to zero, then press the oxygen flush 

button for 5-10 seconds to flush anesthetic from the chamber.

WARNING: ONLY USE THE CHAMBER FLUSH WITH CHAMBERS 

SPECIFICALLY DESIGNED FOR THIS PURPOSE. CHAMBER MUST HAVE A 

POSITIVELY SEALING GASKET BETWEEN THE LID AND BODY OF THE 

CHAMBER.

3.  Quickly open the lid, remove the desired animals and place them on the surgery 

surface.

4.  Remove the black plug from the nosecone and gently slide the animal into the 

nosecone. Be sure that the animal’s face has a tight fit inside the nosecone.

5.  If animals are to remain in the induction chamber, open the secondary flowmeter 

to half of the total inlet flow (For example, 0.5 liters/min. for mice).

6.  If no animals remain in the induction chamber, be sure that the secondary 

flowmeter is turned fully clockwise to the OFF position.

 

NOTE: Use only the EZ-104 Versaflex Breathing Device for animals weighing more 

than 500g. The other units cannot handle the tidal volume of larger animals.

Summary of Contents for EZ Anesthesia

Page 1: ...WORLD PRECISION INSTRUMENTS EZ Anesthesia Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 043019...

Page 2: ......

Page 3: ...ree Phases 3 Anatomy of an Anesthesia System 3 SETUP PROCEDURES 4 Water Pump And Surgical Pump Setup 5 Isoflurane Vaporizer Setup 5 Cautions and Warnings 5 Filling the Vaporizer 6 OPERATING THE SYSTEM...

Page 4: ...ii World Precision Instruments...

Page 5: ...ting instructions and not following standard laboratory safety procedures when working with anesthetic agents and volatile compressed gases The system and its components must only be modified or repai...

Page 6: ...ygen Regulator Flow Meter 1 Vaporizer Manifold Flow Meter 2 Induction Chamber Oxygen Flush Charcoal Filter Canister Charcoal Filter Canister Breathing Device Nose cone Surgical Bed Water Heating Pump...

Page 7: ...g liquid to create an anesthetic gas at the exact concentration level required Manifold distributes anesthetic gas to the induction chamber and the breathing devices Flowmeter 2 controls the flow rate...

Page 8: ...induction chamber input B Attach air filter canister to the white plastic filter adaptor at the end of the large diameter exhaust line coming from the chamber C Slide the air filter canister onto the...

Page 9: ...ls below the indicator line turn off the pump remove cap and add water up to the indicator line Do not exceed the indicator line level Replace the cap and operate the pump Repeat steps 4 and 5 until t...

Page 10: ...vaporizer upright at all times and do not carry the vaporizer by holding the dial control Filling the Vaporizer 1 The control dial must be in OFF position Press the lock button to turn the dial Alway...

Page 11: ...g and surgical bed is at proper temperature WARNING IMPORTANT MAKE SURE THE SMALL BLACK PLUG IS INSERTED INTO BREATHING DEVICE NOSECONE THE OPERATOR COULD FACE EXPOSURE TO ISOFLURANE IF THE PLUG IS NO...

Page 12: ...e moved to the breathing unit be sure that the desired breathing units are connected to the manifold 2 Reduce the secondary flowmeter setting to zero then press the oxygen flush button for 5 10 second...

Page 13: ...o appropriately marked containers and disposed of properly Less frequent intervals may be used when the anesthetic agent does not contain additives or a stabilizing agent CAUTION Do not drain the vapo...

Page 14: ...e top of the water pump and tip the pump over a sink until all the water is emptied from the pump 3 Replace the cap on the pump Disinfecting the Water Pump and Surgical Bed The water pump and surgical...

Page 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Page 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Reviews: