Operating the Device
25 / 62
– The pump can be started/stopped as often as desired with a tube set in-
validated during the current application if the tube set is not removed.
Once the tube set is removed, it can no longer be used.
– If the inserted tube set has already been invalidated, an error message E02
is displayed until the tube set is removed.
– If an invalid tube set is inserted, an error message E03 is displayed until
the tube set is removed.
– If the inserted tube set has been inserted incorrectly and the pump is star-
ted, an error message E04 is displayed for 3 seconds.
– If no tube set has been inserted and the pump is started, an error message
E01 is displayed for 3 seconds.
– If an invalid tube set has been inserted and the pump is started, E03
flashes 3 times.
– If an invalidated tube set has been inserted and the pump is started, E02
flashes 3 times.
• Reusable tube set (max. 20 cycles, NOT FOR SALE IN USA):
Display xxx (3 s) -> remaining cycles after this use
– After the last remaining cycle, 000 is displayed for an additional 3 seconds
with every stop and a low volume acoustic signal (beep) is emitted.
– The tube set is invalidated 1 minute after starting the pump.
– The pump can be started/stopped as often as desired with a tube set in-
validated during the current application if the tube set is not removed.
Once the tube set is removed during the last remaining cycle, this tube set
can no longer be used.
– If the inserted tube set has already been invalidated, an error message E02
is displayed until the tube set is removed.
– If an invalid tube set is inserted, an error message E03 is displayed until
the tube set is removed.
– If the inserted tube set has been inserted incorrectly and the pump is star-
ted, an error message E04 is displayed for 3 seconds.
– If no tube set has been inserted and the pump is started, an error message
E01 is displayed for 3 seconds.
– If an invalid tube set has been inserted and the pump is started, E03
flashes 3 times.
– If an invalidated tube set has been inserted and the pump is started, E02
flashes 3 times.
Transponder signal loss
The device stops and cannot be restarted if the transponder loses its signal during
current operation (e.g. defective electronic component). If the signal is automatic-
ally restored within 1 second, it is possible to continue using the tube set for addi-
tional applications.
5.7
Inserting the Irrigation Tube Sets
The irrigation tube set is available as a disposable or reusable tube set (auto-
clavable, NOT FOR SALE IN USA) (see Chapter Irrigation Tube Sets Overview [
23]).