Operating the Device
23 / 62
Fig. 5–3 Rear of the device
(18)
Service interface
(19)
Toggle screw
(20)
Tripod mount
(21)
Mains socket
(22)
Fuse holder
(20)
(18) (19)
(21)
(22)
5.4
Turning the Device On
1.
Plug the device into the power outlet. The mains voltage LED lights up green
(Front of the device [
(14)
).
2.
Make sure a tube set is not inserted into the tube retainer.
3.
Press "ON" key (Front of the device [
(13)
). The device switches on.
4.
The "ON" operating mode LED lights up green (Front of the device [
(12)
).
A device self-test is carried out automatically.
If a tube set is in the tube retainer of the device when the device is being switched
on, the display for operating states and warnings (Front of the device [
(5)
)
depicts "----" and a warning acoustic signal (beep) is emitted once. The self-test
continues once the tube set is removed.
If the device check is unsuccessful, the corresponding error messages are displayed
on the Operating States and Warnings display ("ERROR" see Chapter Error and
Warning Messages [
54]).
A low volume acoustic signal (beep) indicates the device self-test completed suc-
cessfully.
5.5
Irrigation Tube Sets Overview
A total of 2 different tube sets can be inserted into the tube retainer on the device
side (Device side [
(15)
). The following table lists the usability of each type of
tube set.
Art. No.
Tube description
T0449-01
Disposable tube set, Luer lock connector
and spikes, only inflow
T0454-01
Reusable tube set (NOT FOR SALE IN
USA), Luer lock connector and spikes,
only inflow
WARNING!
Preventing infections
Sterilize reusable instruments and reusable tubes sets before surgery to prevent in-
fections. Check all the single-use/disposable items before removing them from the
package to ensure that the packaging is intact and that the expiration date is still
valid.