Function Test
34 / 62
3.
Start the pump with the START/STOP key. The START operating mode LED (Fig.
Front of the device [
(3)
) lights up. Wait until the instrument recognition
is finished (the instrument is outside of the joint and at a working height of
± 10 cm of the operation height, see Chapter Starting the Device [
29]).
4.
Close the inflow valve after the instrument recognition is finished.
7.3
Checking the WASH Function
1.
Start the Wash function by pressing the WASH key on the device (or the corres-
ponding key on the remote control). The pressure increase for the WASH func-
tion, in this case 150 mmHg, is depicted in the nominal pressure display. The
roller wheel must pump briefly and then stops.
2.
Stop the Wash function by pressing the WASH key on the device once more (or
the corresponding key on the remote control). The pressure display depicts the
last set values, in this case 100 mmHg. The roller wheel must reverse briefly
and then stops.
7.4
Ending the Function Test
Stop the device by pressing the START/STOP key (see Fig. Front of the device [
(2)
).
The function test has now been successfully completed. The device is now ready for
surgical use. Position the device as needed for surgery.