General Information
15 / 62
CAUTION!
Ventilation of the device
• Avoid device overheating.
• Ensure free air circulation especially to the bottom and rear of the device (rear
panel distance of at least 10 cm).
CAUTION!
Patient group
There are no restrictions as to the specification of the patient group when using
the device as intended and its use does not endanger the patient's health.
CAUTION!
Cleaning the Device
Do not sterilize the device.
CAUTION!
Overheating
To avoid overheating,
• the device must not be operated for more than 10 minutes at maximum
speed.
• the device must not be operated without fluid in the tube system.
CAUTION!
Mains Power Cable
Any power cables employed by the user that are not provided by the manufacturer
must meet the safety requirements of the national standards in the respective cur-
rent valid version.
CAUTION!
Device setup
Device should be positioned outside of the sterile area in such a way that
• it can be easily disconnected,
• it is easy to use and switch off and on,
• it allows an easy monitoring of the display values, device functions, and access
to the control elements.
3.4
Device-Specific Warnings: Arthroscopy Indication
WARNING!
Irrigation fluid bags
The device is only intended for use with flexible fluid bags. Do not use glass con-
tainers as they might break. With rigid containers, fluid cannot flow quickly
enough due to the vacuum being generated inside of the containers. Risk of implo-
sion with rigid containers.