9. ADMINISTRATOR MODE
9.6 Table
In this part, you can select statistic for each ROI , extreme and its critical value, operator, digital outputs with
logic and analog outputs.
Each row in the table represents a ROI you have defined before.
9.6.1 Statistic
In each frame there is certain area depending on the size of the ROI. The area consists out of number of pixels,
each with its own temperature. These pixels are used for calculating the statistical value. For each ROI, you can
select:
Max
‐ Maximum value out of the ROI pixels.
Min
‐ Minimum value out of the ROI pixels.
Avr
‐ Average value out of the ROI pixels.
Dev
‐ Standard deviation value out of the ROI pixels.
Var
‐ Variation value out of the ROI pixels.
9.6.2 Value
This represents the current statistical result of the frame for the ROI
9.6.3 Extreme
The statistic value is calculated for each frame. In case of Frame or Start‐Stop mode, the measurement can
consist of several frames and thus several statistical results for each ROI.
The extreme selects either the lowest or the highest value from this sequence of statistical results.
9.6.4 Operator
The one value that comes out of the measurement sequence (or just from one measurement in case of sequence
of one frame or Continuous or Non‐trigger mode) is called measurement value. This value is compared to the
extreme value.
The operator selects, whether the measurement value should be higher (>) or lower (<) than the extreme value.
9.6.5 Extreme val.
Specify the extreme value, that should be compared to the measurement value.
E‐mail and Web
[email protected]
www.workswell.eu
Mobile:
+420 725 877 063
ID:
Reg. No.: 29048575
VAT No.: CZ29048575
Headquarters
Libocka 653/51b
Prague, Czech Republic
Version 210127 EN, 27
th
Jan, 2021
All pictures are only for illustration.
Real values may vary.
31