regularly to ensure cleanliness and prevent the introduction of 

contaminants;

d) Avoid contact with parts of different metals when packaging.

5.8 Sterilization

Use only the following steam sterilization procedures (fractional pre-

vacuum procedure*) for sterilization, and other sterilization procedures 

are prohibited:

1. The steam sterilizer complies with EN13060 or is certified according 

to EN 285 to comply with EN ISO 17665;

2. The highest sterilization temperature is 138 ° C;

3. The sterilization time is at least 4 minutes at a temperature of 

132℃/134℃ and a pressure of 2.0 bar ~ 2.3 bars.

4. Allow a maximum sterilization time of 20 minutes at 134 °C.

Verification of the fundamental suitability of the products for effective 

steam sterilization was provided by a verified testing laboratory.

Notes

a) Only products that have been effectively cleaned and disinfected are 

allowed to be sterilized;

b) Before using the sterilizer for sterilization, read the Instruction Manual 

provided by the equipment manufacturer and follow the instructions.

c) Do not use hot air sterilization and radiation sterilization as this may 

result in damage to the product;

d) Please use the recommended sterilization procedures for sterilization. 

It is not recommended to sterilize with other sterilization procedures 

such as ethylene oxide, formaldehyde and low temperature plasma 

sterilization. The manufacturer assumes no responsibility for the 

procedures that have not been recommended.

If you use the sterilization procedures that have not been recommended, 

please adhere to related effective standards and verify the suitability and 

effectiveness.

* Fractional pre-vacuum procedure = steam sterilization with repetitive 

pre-vacuum. The procedure used here is to perform steam sterilization 

through three pre-vacuums.

Summary of Contents for LED.D

Page 1: ...LED D Curing light Instruction Manual ...

Page 2: ...terilization 3 6 Cautions 9 7 Installation 11 8 Operation 11 9 Maintenance 13 10 Transportation 13 11 Storage and transportation 14 12 Trouble shooting 14 13 Packing list 15 14 Environment protection 15 15 After service 16 16 European authorized representative 16 17 Symbol instructions 16 18 Statement 17 19 EMC Declaration of conformity 18 ...

Page 3: ... a short time 1 2 This product is used for dentistry It has the function of accelerating dental restoration and solidifying the material of dental whitening 2 Structure and components LED D curing light dentistry is mainly composed by high power LED optical fiber and main unit Figure 1 shows the main components of LED D Figure 1 ...

Page 4: ... temperature 5 40 3 4 2 Relative humidity 30 75 3 4 3Atmosphere pressure 70kPa to106kPa 3 5 Size 202mm 32mm 38mm 3 6 Net weight 143g 3 7 Consume power 8W 3 8 Protection type against electrical shock class II 3 9 Protection degree against electrical shock type B 3 10 Protection against harmful ingress of water or particular matter ordinary equipment IPX0 not water protected 3 11 Safety in the prese...

Page 5: ...re selected accordingly However with every renewed preparation for use thermal and chemical stresses will result in ageing of the products The maximum number of sterilizations for optical fiber is 500 times 5 1 Initial processing 5 1 1 Processing principles It is only possible to carry out effective sterilization after the completion of effective cleaning and disinfection Please ensure that as par...

Page 6: ...he optical fiber until the dirt on surface is not visible Then use soft cloth to dry the optical fiber and put them into a clean tray The cleaning agent can be pure water distilled water or deionized water 5 3 Cleaning The cleaning should be performed no later than 24 hours after the operation The cleaning can be divided into automated cleaning and manual cleaning Automated cleaning is preferred i...

Page 7: ...rature disinfection function The temperature does not exceed 134 and the disinfection under the temperature cannot exceed 20 minutes The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883 Cleaning and disinfecting steps by using Washer disinfector 1 Carefully place the product into the disinfection basket Fixation of product is needed only when the product is removable...

Page 8: ...entration and time provided by manufacturer shall be obeyed The used cleaner is neodisher MediZym Dr Weigert d Disinfection d1 Direct use after disinfection temperature 90 C time 5 min or A0 3000 d2 Sterilize it after disinfection and use temperature 90 C time 1 min or A0 600 d3 For the disinfection here the temperature is 93 C the time is 2 5 min and A0 3000 e Only distilled or deionized water wi...

Page 9: ...the product If there is still visible stain on the product after cleaning disinfection the entire cleaning disinfection process must be repeated 5 6 2 Check the product If it is obviously damaged smashed detached corroded or bent it must be scrapped and not allowed to continue to be used 5 6 3 Check the product If the accessories are found to be damaged please replace it before use And the new acc...

Page 10: ...es a Only products that have been effectively cleaned and disinfected are allowed to be sterilized b Before using the sterilizer for sterilization read the Instruction Manual provided by the equipment manufacturer and follow the instructions c Do not use hot air sterilization and radiation sterilization as this may result in damage to the product d Please use the recommended sterilization procedur...

Page 11: ...e 2 It should not be mixed with dangerous goodsduring transportation 3 Avoid exposure to sun or rain or snow during transportation The cleaning and disinfection of main unit are as follows Before each use wipe the surface of the machine with a soft cloth or paper towel soaked in 75 medical alcohol Repeat the wipe for at least 3 times After each use wipe the surface of the device with a soft cloth ...

Page 12: ...serve separately 6 11 This equipment may cause the electromagnetic interference Please don t use it on the patients with pacemaker or on an E Surgery Please be caution to use this equipment under the condition of strong electromagnetic interference 6 12 It is forbidden to disassemble the equipment We shall not assume any responsibility for any malfunction damage or accident caused by disassembling...

Page 13: ...of the new battery correctly Figure 2 put the new battery in and then fix the battery cover 7 4 Charge method For Charging please pull out the charge port protector of the main unit slightly plug the output port of the adapter into the charge port of the main unit and then connect the adapter to the power supply When charging is finished pull out the adapter and plug the protector back into the ch...

Page 14: ...ed inside the main unit when low power is detected the indicator of the main unit will wink please charge in time 8 6 The mode indictor will be enlightened one by one when charging They will all be enlightened if charging has finished If a battery error is detected the indicator of the mode will wink 8 7 When the equipment is charging it is possible to turn the equipment into the normal work state...

Page 15: ...on then it will come into over heat protection status And only after 2 minute sleep it can restart working 200s continuously 9 Maintenance 9 1 This equipment does not include the self maintenance parts so it should be performed by professional or special maintenance shop 9 2 The optical fiber should be sterilized for 4 minutes with 134 and 2 0bar 2 3bar 0 20MPa 0 23MPa before each use other parts ...

Page 16: ...nt where the humidity is 10 93 the atmosphere pressure is 70kPa 106kPa and the temperature is 20 55 11 4 Excess impact or shake should be prevented during transportation Handle with care Do not place upside down 11 5 Don t put it together with dangerous articles during transportation 11 6 Keep it away from the sun rain or snow during transportation 12 Trouble shooting Faulty Possible cause Solutio...

Page 17: ... 3 Cleaned by the alcoho Light intensity is weak 1 The optical fiber is not installed well 2 There is crevice on the optical fiber 3 There is resin on the tip of the optical fiber 1 Reinstall the optical fiber 2 Change a new optical fiber 3 Clear the resin Effective duration of the battery become short The capacity of the battery decreased Change a new battery T h e m o d e i n d i c a t o r twink...

Page 18: ...ir to the equipment if it has quality problems please refer to the warranty card for the warranty period 16 European authorized representative 17 Symbol instructions Trademark CE marked product Type B applied part Screw inside outside Ordinary equipment Class II equipment Date of manufacture Alternating current Manufacturer Recovery ...

Page 19: ... Atmospheric pressure for storage Appliance compliance WEEE directive Follow Instructions for Use Authorised Representative in the EUROPEAN COMMUNITY 18 Statement All rights of modifying the product are reserved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to GUILIN WOODPECKER MEDICAL ...

Page 20: ...ake product must take legal responsibilities 19 EMC Declaration of conformity The device has been tested and homologated in accordance with EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment ...

Page 21: ...19 ...

Page 22: ...20 ...

Page 23: ...21 ...

Page 24: ...ZMN SM 060 V4 2 2020628 ...

Reviews: